Nikon ECLIPSE TE2000-E Instructions Manual page 4

Inverted microscope
Table of Contents

Advertisement

1.
Intended use of the equipment.
This microscope is intended mainly for use in microscopic observation and in the micromanipulation of
living cells and tissue using diascopic (transmitted) and episcopic (reflected) illumination. It is designed
for the main purposes of experimentation and observation, in hospitals or other laboratories, of such cells
and tissue within the fields of genetics, immunology, physiology, pharmacology, neurology, cellular biology,
and molecular biology.
2.
Do not disassemble.
Disassembly may cause malfunction and / or electrical shock, and will lead to the forfeiture of all claims
against warranty.
experience any problem with the microscope, notify your nearest Nikon representative.
3.
Check the Input Voltage.
This microscope uses a power supply for the lamp.
When using the power supply TE-PS30 or TE-PSE30:
Make sure that the input voltage indication on the rear panel matches your regional voltage supply. If the
voltages do not match, do not use the microscope; instead, notify your nearest Nikon representative.
Using the power supply with the wrong input voltage may cause a short circuit or fire. It may also cause
damage to the microscope.
When using the power supply TE2-PS100W:
If you are using the power supply TE2-PS100W, there is no need to check the input voltage since the input
voltage of this power supply is AC 100-240V and can be used at any place in the world.
4.
Check the AC adapter of the HUB controller (when using the T-HUBC HUB controller).
The HUB controller is powered by the AC adapter. Be sure to use the specified adapter model meeting the
requirements given below. Use of any other type of adapter can result in malfunction, excessive heating,
and/or fire.
To prevent malfunction and/or fire, be sure to use the AC adapter in a well-ventilated location.
ensure that it radiates heat properly and does not overheat, never cover or place any object on the
adapter.
To prevent malfunction, always turn off the power switch (switch to "O") of the HUB controller before
attaching the AC adapter.
Specified AC adapter
Manufacturer:
Model:
Rated input voltage: AC 100-240 V, 0.7 A, 50/60 Hz
Voltage fluctuation:
Rated output voltage: DC 12 V
Rated output current: 2.5 A
Others:
5.
Power cord for power supply and the power cord for AC adaptor of the HUB controller
To prevent electric shock, always turn off the power switch (switch to "O") for the power supply and the HUB
controller before attaching or detaching the power cord. Use one of the power cords specified below. Use of an
improper power cord can result in fire or other hazard. Also note that the power supply is classified as subject
to protection class I against electric shock. Therefore, be sure to connect it to a protective ground terminal.
Using units in areas where the supply voltage is 100 to 120 V
UL Listed detachable power cord set, 3 conductor grounding Type SVT, No. 18 AWG, 3 m long maximum,
rated at 125 V AC minimum.
Using units in areas where the supply voltage is 220 to 240 V
Approved according to EU/EN standards, 3 conductor grounding Type HO5VV-F, 3 m long maximum,
rated at 250 V AC minimum.
6.
Heat from the light source.
The lamp and the lamphouse become extremely hot by the lamp illumination. To avoid burns, do not
touch the lamphouse while the lamp is lit or for thirty minutes after it has been turned off.
To avoid the risk of fire, do not place fabric, paper or highly flammable volatile materials such as
gasoline, petroleum benzine, paint thinner or alcohol near the lamphouse while the lamp is lit or for
about thirty minutes after it has been turned off.
The bottom plate of the power supply becomes hot during use. Do not cover up the ventilation holes
on the side of the power supply.
2
Do not disassemble any part other than those described in this manual. If you
PHIHONG ENTERPRISE (Taiwan)
PSA30U-120 (N)
±10%
UL Listed product, GS approved, CE satisfied
To