Nikon ECLIPSE TE2000-E Instructions Manual page 23

Inverted microscope
Table of Contents

Advertisement

2.
Turn on the dia-illumination.
1) Set the "EXTERNAL switch" on the rear side of the power
supply to ON.
2) Turn on the "power switch" of the power supply. (Flip the
switch to .)
3) Press the "dia-illumination ON/OFF switch" on the left side
of the microscope to turn on the lamp.
3.
Adjust the lamp to ensure color fidelity.
1) Set the "brightness adjustment dial" on the left side of the
microscope to the < 12V100W > indication.
2) Move the "NCB11 filter" on the dia-illuminator into the
optical path.
3) Move the "ND4 filter" on the dia-illuminator into the optical
path.
4.
Adjust the optical path.
1) Move the < 10X > objective into the optical path.
2) Set the "optical path switchover dial" on the right side of the
microscope to position < 1 >. (-> Direct 100% light to the
observation port.)
3) Raise the "field diaphragm lever" on the dia-illuminator
completely to open the field diaphragm fully.
4) Turn the "aperture diaphragm lever" on the system condenser
to the right limit to open the "aperture diaphragm" fully.
5) Turn the condenser focus knob to lower the condenser
mount to the lowest position.
Supplementary information
When the system condenser is mounted with the "ELWD
condenser lens", position the condenser mount at the
location about 1 cm below the uppermost position.
When using the "ELWD-S condenser", position the
condenser mount at the location about 2 cm below the
uppermost position.
5.
Set the microscope for bright-field microscopy.
1) Set the "condenser turret" to position < A >.
6.
Adjust the diopter and interpupillary distance.
1) Turn the "photo-mask dial" on the front side of the
microscope counterclockwise to move the photo-mask into
the optical path.
2) Look into the left eyepiece with your left eye. Turn the
"diopter adjustment ring" of the left eyepiece to bring the
double crosshairs of the photo-mask into focus.
3) Look into the right eyepiece with your right eye. Turn the
"diopter adjustment ring" of the right eyepiece to bring the
double crosshairs of the photo-mask into focus.
II. Microscopy
Turn the diopter adjustment ring
so that the double crosshairs come into focus.
Adjust the interpupillary distance
to consolidate the view fields of both eyepieces.
21