Time Required For Calibration; Automating Calibration Procedures - Agilent Technologies 34410A Service Manual

6 1/2 digit multimeter
Hide thumbs Also See for 34410A:
Table of Contents

Advertisement

Time Required for Calibration

The instrument can be automatically calibrated under computer control. With
computer control you can perform the complete calibration procedure and
performance verification tests in less than 30 minutes once the instrument is
warmed–up (see Test Considerations on

Automating Calibration Procedures

The adjustment procedures provided in this Service Guide demonstrate
front panel adjustment. You can automate the complete verification and
adjustment procedures outlined in this manual. You can program the
instrument configurations specified for each test over the remote interface.
You can then enter read back verification data into a test program and
compare the results to the appropriate test limit values.
The instrument must be unsecured prior to initiating the calibration
procedure (see "Calibration Security" on
Refer to the Agilent 34410A/11A/L4411A Programmer's Reference Help for
information on SCPI programming for the multimeter. This is a standard
Windows online help system, provided with the multimeter on the Agilent
34410A/11A/L4411A Product Reference CD–ROM. It is also downloadable
from the Web. See either product page:
34410A/11A/L4411A Service Guide
www.agilent.com/find/34410A
www.agilent.com/find/34411A
www.agilent.com/find/L4411A
Calibration Procedures
page
63).
page
75).
3
59

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34411a

Table of Contents