Agilent Technologies 3458A Datasheet

Hide thumbs Also See for 3458A:

Advertisement

Shattering
performance
barriers
of speed
and
accuracy
Agilent 3458A Multimeter
Data Sheet

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 3458A

  • Page 1 Agilent 3458A Multimeter Data Sheet Shattering performance barriers of speed accuracy...
  • Page 2 Performance Highlights dc volts ac volts Frequency and period • 5 ranges: 0.1 V to 1000 V • 6 ranges: 10 mV to 1000 V • Voltage or current ranges • 8.5 to 4.5 digit resolution • 1 Hz to 10 MHz bandwidth •...
  • Page 3 Access speed and accuracy through a powerful, convenient front panel Display Numeric/user keys Current measurement terminals • Bright, easy-to-read, vacuum • Numeric entry for constants and • Easy fuse replacement with fuse fluorescent display measurement parameters holder built into terminal •...
  • Page 4: Table Of Contents

    The 3458A is simply the fastest, most flexible, and most accurate multimeter ever offered by Agilent Technologies. In your system or on the bench, the 3458A saves you time and money with unprecedented test system throughput and accuracy, seven function measurement flexibility, and low cost of ownership.
  • Page 5: High Resolution Digitizing /8

    The 3458A Multimeter for: High test system throughput Faster testing • Up to 100,000 readings/sec • Internal test setups > 340/sec • Programmable integration times from 500 ns to 1 sec Greater test yield • More accuracy for tighter test margins •...
  • Page 6: Test System Throughput

    The 3458A Multimeter fits fast and accurate measurements. additional years of return-to-repair. your needs from 4.5 digit dc volts Consider the 3458A’s 0.6 ppm 24 hour measurements at 100,000/second, dc volts accuracy, 100 ppm ac volts to 8.5 digit dc volts measurements at...
  • Page 7: Calibration Lab Precision

    • 0.1 ppm dcV linearity • 100 ppm acV absolute accuracy • 4 ppm/year optional stability In the calibration lab, you’ll find the Finally, the 3458A, like its dmm pre- Easy calibration 3458A’s 8.5 digits to have extraordi- decessors, offers offset-compensated...
  • Page 8 The maximum sample rate trigger. Once synchronized, the dmm is 50,000 samples/second or 20 µs automatically acquires the waveform The 3458A gives you the choice of between samples. Samples can be through digitizing successive periods two configurations for high speed internally paced by a 0.01% accurate...
  • Page 9: Technical Specifications

    3458A Technical Specifications...
  • Page 10: Section 1: Dc Voltage

    Section 1: DC Voltage...
  • Page 11: Section 2: Resistance

    DC Voltage continued Section 2: Resistance...
  • Page 12 Resistance continued...
  • Page 13: Section 3: Dc Current

    Section 3: DC Current...
  • Page 14: Section 4: Ac Voltage

    Section 4: AC Voltage General Information Synchronous Sub-sampled Mode (ACV Function, SETACV SYNC)
  • Page 15 AC Voltage continued...
  • Page 16 AC Voltage continued Analog Mode (ACV Function, SETACV ANA)
  • Page 17 AC Voltage continued Random Sampled Mode (ACV Function, SETACV RNDM)
  • Page 18 AC Voltage continued...
  • Page 19: Section 5: Ac Current

    Section 5: AC Current...
  • Page 20: Section 6: Frequency/Period

    AC Current continued Section 6: Frequency/Period...
  • Page 21: Section 7: Digitizing

    Section 7: Digitizing Standard DC Volts Digitizing (DCV Function)
  • Page 22 Digitizing continued Direct and Sub-sampled Digitizing (DSDC, DSAC, SSDC and SSAC Functions)
  • Page 23: Section 8: System Specifications

    Section 8: System Specifications...
  • Page 24: Section 9: Ratio

    Section 9: Ratio Section 10: Math Functions...
  • Page 25: Section 11: General Specifications

    Input terminals Gold-plated tellurium copper IEEE-4888 interface Complies with the following: Included with 3458A IEEE-488.1 Interface Standard 34137A Test lead set for 3458A IEEE-728 Codes/Formats Standard calibration certificate HPML (Multimeter Language) Storage temperature –40° C to +75° C Warm-up time...
  • Page 26: Section 12: Ordering Information

    11059A Kelvin probe set (4-wires, 1 m) 11060A Surface mount device (SMD) test probes 11062A Kelvin clip set (2 each) 34137A Deluxe test lead set for the 3458A 34308A Thermistor kit 34330A 30 A current shunt E2308A Thermistor temperature probe...
  • Page 27 Ethernet and Taiwan 0800 047 866 the Web inside your test systems. www.agilent.com/quality Other AP Countries (65) 375 8100 Agilent is a founding member of the LXI consortium. Europe & Middle East Belgium 32 (0) 2 404 93 40 Denmark...