Optional Ac Voltage Performance Verification Tests - Agilent Technologies 34410A Service Manual

6 1/2 digit multimeter
Hide thumbs Also See for 34410A:
Table of Contents

Advertisement

3
Calibration Procedures

Optional AC Voltage Performance Verification Tests

Configuration:
1 Make sure you have read
2 Connect the calibrator to the input terminals.
3 Set the AC Volts function and the 3 Hz input filter. With the slow filter
4 Select each range in the order shown below. Provide the indicated input
72
AC Volts (CONFigure[:VOLTage]:AC)
LF 3 HZ:SLOW ([SENSe:]VOLTage:AC:BANDwidth 3)
a For the 34410A/11A use the front panel input terminals and select
the Front input terminals with the Front/Rear switch.
selected, each measurement takes 2.5 seconds to complete.
voltage and frequency. Compare measurement results to the appropriate
test limits shown in the table. (Be certain to allow for appropriate
source settling.)
Input
Vrms
Frequency
1 V
10 Hz
1 V
1 kHz
1 V
20 kHz
1 V
50 kHz
1 V
100 kHz
1 V
300 kHz
10 V
1 kHz
1 V
1 kHz
0.1 V
1 kHz
"Test Considerations"
Error from Nominal
Range
24 hour
1 V
± 430 µV
1 V
± 400 µV
1 V
± 400 µV
1 V
± 900 µV
1 V
± 2.8 mV
1 V
± 15 mV
10 V
± 4 mV
10 V
± 2.2 mV
10 V
± 2.02 mV
34410A/11A/L4411A Service Guide
on page 63.
90 day
1 year
± 830 µV
± 930 µV
± 800 µV
± 900 µV
± 800 µV
± 900 µV
± 1.4 mV
± 1.5 mV
± 3.8 mV
± 4.8 mV
± 17 mV
± 17 mV
± 8 mV
± 9 mV
± 3.5 mV
± 3.6 mV
± 3.05 mV
± 3.06 mV