Saving and Printing Data
N O T E
248
of the 1000 horizontal data points, and min and max values will
be displayed on screen. However, the oversampled data is also
processed to provide a best-estimate value at each of the 1000
horizontal points. The data in the CSV file will be the
best-estimate value at each of the 1000 horizontal points.
Therefore, the min and max values may not appear in the CSV
file.
This occurs when oversampling occurs
(10 * seconds per division * maximum sample rate > 1000).
To save the CSV, ASCII XY, or BIN data, or BMP or PNG images to the USB
storage device, it is necessary to press the Quick Print button (see
print the display"
on page 66).
6000 Series Oscilloscope User's Guide
"To