Agilent Technologies 6000 Series User Manual page 106

Oscilloscope
Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

4
Triggering the Oscilloscope
106
Adjust the trigger level for the selected analog channel by
turning the Trigger Level knob. Press the D15 Thru D0 key and
select Thresholds to set the threshold level for digital
channels. The value of the trigger level or digital threshold is
displayed in the upper-right corner of the display.
3 For each channel you select, press one of the condition
softkeys to set the condition for that channel in the pattern.
H sets the pattern to high on the selected channel. A high
is a voltage level that is greater than the channel's trigger
level or threshold level.
L sets the pattern to low on the selected channel. A low is a
voltage level that is less than the channel's trigger level or
threshold level.
X sets the pattern to don't care on the selected channel.
Any channel set to don't care is ignored and is not used as
part of the pattern. However, if all channels in the pattern
are set to don't care, the oscilloscope will not trigger.
The rising edge ( ) or falling edge ( ) falling edge softkey
sets the pattern to an edge on the selected channel. Only
one rising or falling edge can be specified in the pattern.
When an edge is specified, the oscilloscope will trigger at
the edge specified if the pattern set for the other channels
is true.
If no edge is specified, the oscilloscope will trigger on the
last edge that makes the pattern true.
Specifying an Edge in a Pattern
You are allowed to specify only one rising or falling edge term in the pattern. If you
define an edge term, then select a different channel in the pattern and define
another edge term, the previous edge definition is changed to a don't care.
6000 Series Oscilloscope User's Guide