Triggering the Oscilloscope
124
4 Press the Signals softkey. The LIN Signals Menu will be
displayed.
Source
Signal
Menu
baud rate
5 Press the Source softkey to select the channel connected to
the LIN signal line.
Adjust the trigger level for the selected analog channel by
turning the Trigger Level knob. Press the D15 Thru D0 key and
select Thresholds to set the threshold level for digital
channels. The value of the trigger level or digital threshold is
displayed in the upper-right corner of the display.
6 Press the Baud softkey to set the LIN signal baud rate to
match your LIN bus signal.
The LIN baud rate can be set to 2400 b/s, 9600 b/s, or 19.2
kb/s. The default baud rate is 2400 b/s.
7 Press the Smpl Pt softkey to select the sample point at which
the oscilloscope will sample the bit value.
Sample Point
8 Press the Standard softkey to select the LIN standard your are
measuring (LIN 1.3 or LIN 2.0).
9 Press the Sync Break softkey and select the minimum number
of clocks that define a sync break in your LIN signal.
Sample
Standard
Point
Selector
One Bit
60%
70%
80%
6000 Series Oscilloscope User's Guide