ARM

Subsystem Syntax

ARM:COUNt

Parameters
Comments
Example
Chapter 3
The ARM subsystem selects the number of scanning cycles (1 to 32,767)
for each INITiate command.
ARM
:COUNt <number> MIN | MAX
:COUNt? [<MIN | MAX>]
ARM:COUNt <number> MIN | MAX
times (1 to 32,767) with one INITiate command when INITiate:CONTinuous
OFF | 0 is set. MIN sets 1 cycle and MAX sets 32,767 cycles.
Name
Type
<number>
numeric
Use only numeric values between 1 and 32767, MIN, or
Number of Scans:
MAX for the number of scanning cycles.
ABORt, INITiate[:IMMediate]
Related Commands:
ARM:COUNt 1
*RST Condition:
Setting Ten Scanning Cycles
This example sets a Form C switch for 10 scans of channels 00 through 03.
When the scan sequence completes, channels 00 through 03 (relays 00
through 03) are closed.
ARM:COUN 10
SCAN(@100:103)
INIT
allows scanning to occur a multiple of
Range of Values
1 - 32,767 | MIN | MAX
!Set 10 scans per INIT command
!Scan channels 00 to 03
!Start scan, close channel 00
E1463A Command Reference 51
Default Value
1