Output[:State]; Output:ttltrgn[:State] - Agilent Technologies E1463A User's Manual And Scpi Programming Manual

32-channel, 5 amp, form c switch scpi programming guide
Hide thumbs Also See for E1463A:
Table of Contents

Advertisement

OUTPut[:STATe]?

Example

OUTPut:TTLTrgn[:STATe]

Parameters
Comments
Chapter 3
queries the present state of the E1406A Command
OUTPut[:STATe]?
Module "Trig Out" port. The command returns "1" if the port is enabled or
"0" if the port is disabled. This command functions the same as
OUTPut:EXTernal[:STATe]?.
Query "Trig Out" Port Enable State
This example enables the E1406A Command Module "Trig Out" port and
queries the enable state. OUTPut[:STATe]? returns "1" since the port is
enabled.
OUTP ON
OUTP?
OUTPut:TTLTrgn[:STATe] <mode>
bus line (0 to 7) will output a trigger when a channel is closed during a scan.
This is also used to disable a selected TTL Trigger bus line. "n" specifies the
TTL Trigger bus line (0 to 7) and <mode> enables (ON or 1) or disables
(OFF or 0) the specified TTL Trigger bus line.
Name
Type
n
numeric
<mode>
boolean
Enabling TTL Trigger Bus:
TTL Trigger bus line (0 to 7) after each channel in the switchbox is closed
during a scan. If disabled, a pulse is not output. The output is a
negative-going pulse.
One Output Selected at a Time:
enabled at one time. Enabling a different output source will automatically
disable the active output. For example, if TTLTrg1 is the active output and
TTLTrg4 is enabled, TTLTrg1 will become disabled and TTLTrg4 will
become the active output.
[ROUTe:]SCAN, TRIGger:SOURce,
Related Commands:
OUTPut:TTLTrgn[:STATe]?
OUTPut:TTLTrgn[:STATe] OFF (disabled)
*RST Condition:
!Enable "Trig Out" port
!Query port enable state
selects and enables which TTL Trigger
Range of Values
0 to 7
ON | OFF | 1 | 0
When enabled, a pulse is output from the selected
Only one output (TTLTrg or EXTernal) can be
E1463A Command Reference 59
Default Value
N/A
OFF | 0

Advertisement

Table of Contents
loading

Table of Contents