OUTPut

Subsystem Syntax

OUTPut:EXTernal[:STATe]

Parameters
Comments
Example
Chapter 3
The OUTPut command subsystem enables or disables the different trigger
lines of the E1406A Command Module.
OUTPut
:EXTernal
[:STATe] <mode>
[:STATe]?
[:STATe] <mode>
[:STATe]?
:TTLTrgn (:TTLTrg0 through :TTLTrg7)
[:STATe] <mode>
[:STATe]?
OUTPut:EXTernal[:STATe] <mode>
the E1406A Command Module.
Name
Type
<mode>
boolean
Enabling "Trig Out" Port:
port after each scanned switchbox channel is closed. If disabled, a pulse is
not output from the port after channel closures. The output pulse is a +5V
negative-going pulse.
"Trig Out" Port Shared by Switchboxes:
pulsed by any switchbox each time a scanned channel is closed. To disable
the output for a specific module send OUTPut:EXTernal[:STATe] OFF or
OUTPut:EXTernal[:STATe] 0 for that module.
One Output Selected at a Time:
enabled at one time. Enabling a different output source will automatically
disable the active output.
[ROUTe:]SCAN, TRIGger:SOURce
Related Commands:
OUTPut:EXTernal[:STATe] OFF (port disabled)
*RST Condition:
Enabling "Trig Out" Port
OUTP:EXT ON
enables or disables the "Trig Out" port on
Range of Values
ON | OFF | 1 | 0
When enabled, a pulse is output from the "Trig Out"
When enabled, the "Trig Out" port is
Only one output (TTLTrg or EXTernal) can be
!Enable "Trig Out" port to output
pulse after each scanned channel
is closed
E1463A Command Reference 57
Default Value
OFF | 0