Agilent Technologies E1463A User's Manual And Scpi Programming Manual page 12

32-channel, 5 amp, form c switch scpi programming guide
Hide thumbs Also See for E1463A:
Table of Contents

Advertisement

Channel 0
Channel 15
Channel 16
Channel 31
12 Getting Started
E1463A MODULE
NO
NC
C0N0
C0C
C0NC
NO
NC
Pads
for MOVS
C15N0
C15C
C15C
for MOVS
NO
NC
C16N0
C16C
C16C
NO
NC
C31N0
C31C
C31C
Figure 1-1. Form C Switch Simplified Schematic
Each channel is switched by opening or closing the appropriate channel
relay. Since the relays are nonlatching, the relays are all open during
power-up or power-down.
When a reset occurs, all channel commons (C) are connected to the
corresponding normally closed (NC) contacts. When a channel is closed,
the common contact (C) is connected to the normally open contact (NO).
User inputs and outputs to each channel are via the NO, NC, and C terminal
connectors on the terminal module.
TERMINAL MODULE
C0NO
C0C
C0NC
C15NO
Jumper
C15C
C15NC
Pads
C16NO
C16C
C16NC
C31NO
C31C
C31NC