Status:operation[:Event]; Status:preset - Agilent Technologies E1463A User's Manual And Scpi Programming Manual

32-channel, 5 amp, form c switch scpi programming guide
Hide thumbs Also See for E1463A:
Table of Contents

Advertisement

Example

STATus:OPERation[:EVENt]?

Comments
Example

STATus:PRESet

68 E1463A Command Reference
Querying the Enable Register in the OPERation Status Register
STAT:OPER:ENAB?
STATus:OPERation[:EVENt]?
the OPERation Status Register are set. The Event Register indicates that
a time-related instrument event has occurred.
Setting Bit 8 of the OPERation Status Register:
to 1 after a scanning cycle completes. Bit 8 returns to 0 (zero) after sending
STATus:OPERation[:EVENt]?.
Returned Data after sending STATus:OPERation[:EVENt]?:
returns "+256" if bit 8 of the OPERation Status Register is set to 1. The
command returns "+0" if bit 8 of the OPERation Status Register is set to 0.
Event Register Cleared:
Status Register with STATus:OPERation:EVENt? clears the Event Register.
Aborting a scan will leave bit 8 set to 0.
Aborting a Scan:
[ROUTe:]SCAN
Related Commands:
Reading the OPERation Status Register After a Scanning Cycle
STAT:OPER?
read the register value
affects only the Enable Register within the OPERation
STATus:PRESet
Status Register by setting all Enable Register bits to 0. It does not affect
either the Status Byte Register or the Standard Event Status Register.
STATus:PRESet does not clear any of the Event Registers.
!Query the Enable Register in the
OPERation Status Register
returns which bits in the Event Register within
Bit 8 (Scan Complete) is set
Reading the Event Register within the OPERation
!Return the bit values of the Event
Register within the OPERation
Status Register
+256 shows bit 8 is set to 1.
+0 shows bit 8 is set to 0.