[ROUTe:]

Subsystem Syntax

[ROUTe:]CLOSe

Parameters
Comments
Chapter 3
The [ROUTe:] command subsystem controls switching and scanning
operations for Form C switch modules in a switchbox.
[ROUTe:]
CLOSe <channel_list>
CLOSe? <channel_list>
OPEN <channel_list>
OPEN? <channel_list>
SCAN <channel_list>
NOTE
There must be a space between the second level command (CLOS, for
example) and the parameter <channel_list>.
[ROUTe:]CLOSe <channel_list>
by <channel_list>. <channel_list> has the form (@ccnn) where cc = card
number (01-99) and nn = channel number (00-31).
Name
<channel_list>
Closing Channels:
To close a single channel use ROUT:CLOS (@ccnn)
To close multiple channels use ROUT:CLOS (@ccnn,ccnn,...)
To close sequential channels use ROUT:CLOS (@ccnn:ccnn)
To close groups of sequential channels use ROUT:CLOS
(@ccnn:ccnn,ccnn:ccnn)
or any combination of the above
NOTE
Closure order for multiple channels with a single command is not
guaranteed. Channel numbers can be in the <channel_list> in any
random order.
Related Commands:
*RST Condition:
closes the Form C switch channels specified
Type
Range of Values
numeric
[ROUTe:]OPEN, [ROUTe:]CLOSe?
All channels open.
E1463A Command Reference 61
Default Value
cc00 - cc31
N/A