3 Installation
Figure 3-1 is a view with the cover removed showing the external Gas
Connector Panel and the internal Electrical Connector Panel.
Figure 3-1: View of Analysis Unit Showing Connector Panels
3.3

Gas Connector Panel Connections

Before using this instrument, it should be determined if the unit will be
used for pressurized service or vacuum service and low pressure applica-
tions. Inspect the restrictor kit that came with the unit. The kit consist of two
restrictors and a union for 1/4" diameter tubing. Notice that the two 1 3/4"
long, 1/4" diameter tubing are restrictors. It has an open end and a closed
end with a small circular orifice. The restrictor without the blue sticker is for
low pressure and vacuum service. For high pressure (5 to 50 psig) applica-
tions, use the restrictor that has a blue sticker on the body.
3-2: Part II