Rear Panel Connections - Teledyne 3000PA-EU Operating Instructions Manual

Percent oxygen analyzer
Table of Contents

Advertisement

3 Installation
3 Installation
3 Installation
3 Installation
3 Installation
Figure 3-1: Front Panel of the Model 3000PA
All operator controls are mounted on the control panel, which is hinged
on the left edge and doubles as the door that provides access to the sensor
and cell block inside the instrument. The door is spring loaded and will
swing open when the button in the center of the latch (upper right corner) is
pressed all the way in with a narrow gauge tool (less than 0.18 inch wide),
such as a small hex wrench or screwdriver Allow clearance for the door to
open in a 90-degree arc of radius 7.125 inches. See Figure 3-2.
Figure 3-2: Required Front Door Clearance
3.3

Rear Panel Connections

Figure 3-3 shows the Model 3000PA rear panel. It contains all of the
gas and electrical inputs and outputs. Some ports are optional equipment.
Refer to page iii in the front of this manual for options included in your
instrument. Be sure to note the instrument serial number.
3-2
Teledyne Analytical Instruments
Model 3000PA