Cell Warranty; Fuse Replacement - Teledyne 3010TA Operating Instructions Manual

Trace oxygen analyzer
Table of Contents

Advertisement

5
Maintenance
Step 9. Push the gate on the cell block down so that the slots on the
side of the gate engage the locating screws on the side of the
block. This forces the holder into position and forms a gas-
tight seal.
Step 10. Purge system with sample or zero gas.
Step 11. Power-up.
If steps 4 through 10 are accomplished quickly (elapsed time less than 15
seconds), recovery to less than 1ppm level should occur in less than 8
hours.

5.2.5 Cell Warranty

The Micro-Fuel cell used in the standard Model 3010T is the L-2C for
Trace Analysis. Check Specific Model Information in the front matter of this
manual for cell class in your unit, if nonstandard, as this will affect cell life
and warranty data. Also note any Addenda that may be attached to the front
of this manual for special information applying to your instrument.
With regard to spare cells, warranty period begins on the date of ship-
ment. The customer should purchase only one spare cell. Do not attempt to
stockpile spare cells.
Note: The L-2C cell is not designed for applications where CO
major component in the sample, however concentrations of
1,000 ppm or less will not adversely effect the cell
performance. Consult TETAI for available options for either
intermittent or continuous CO
If a cell was working satisfactorily, but ceases to function before the
warranty period expires, the customer will receive credit toward the purchase
of a new cell.
If you have a warranty claim, you must return the cell in question to the
factory for evaluation. If it is determined that failure is due to faulty work-
manship or material, the cell will be replaced at no cost to you.
Note: Evidence of damage due to tampering or mishandling will
render the cell warranty null and void.
5.3

Fuse Replacement

The 3010T Analysis Unit requires two 5 x 20 mm, 1.6 A, T type (Slow
Blow) fuses. The fuses are located inside the explosion proof housing on the
Electrical Connector Panel, as shown in Figure 5-3. To replace a fuse:
5-6: Part II