Ac Power; Connecting The Instrument To A Power Source - Fluke NetDAQ 2640A User Manual

Networked data acquisition unit
Hide thumbs Also See for NetDAQ 2640A:
Table of Contents

Advertisement

2640A/2645A NetDAQ
Users Manual
107-264V
50/60 Hz
15VA
ALARM/TRIGGER I/O
+
9-16V
DC PWR
+ –
External Battery (DC Operation)
WARNING:
IF VOLTAGES GREATER THAN 50V (2640A ONLY) ARE TO BE MEASURED, A
SEPARATE EARTH GROUND MUST BE ATTACHED TO THIS REAR PANEL
GROUND CONNECTOR WHEN THE INSTRUMENT IS OPERATED FROM BATTERY
POWER.
To avoid shock hazard when powering the instrument
with ac power, connect the power cord to a receptacle
with an earth ground.

AC Power

Plug the line cord into the connector on the rear of the instrument as shown in
Figure 2-2. The instrument operates on any line voltage between 107 and 264V ac
(45 to 65 Hz) without adjustment. Fluke warrants the instrument to meet
specifications only at 50 Hz and 60 Hz operation. Power consumption is a
nominal 15 watts. Be sure the line cord ground terminal connects to an earth
ground.
To avoid shock hazard when powering the instrument
with dc power, connect the instrument ground terminal to
an earth ground.
2-6
Line Cord (AC Operation)
OVERVOLTAGE CATEGORY II PER IEC 1010-1
ON / OFF
DIGITAL I/O
MA TO TI
0 1 2
3 4 5 6 7
+30V
MEETS VFG 243 / 1991
Figure 2-2. Connecting the Instrument to a Power Source
P
P
MODEL
: 2640A / 41A
WARNING
: TO AVOID ELECTRICAL SHOCK, DISCONNECT LINE CORD BEFORE REMOVING COVER
SERIAL PORT
Ground for 50-ohm
Termination Ground Lug.
Warning
Warning
2645A / 46A
XMT RCV
NOT FOR CONNECTION TO
PUBLIC TELEPHONE SYSTEMS
ETHERNET
LK
2-6.