A/2645A Rear Panel - Fluke NetDAQ 2640A User Manual

Networked data acquisition unit
Hide thumbs Also See for NetDAQ 2640A:
Table of Contents

Advertisement

2640A/2645A NetDAQ
Users Manual
Ground Terminal.
Connects mainframe to ground.
Power Switch.
Applies power to the
instrument (ac or dc
operation).
107-264V
50/60 Hz
15VA
ALARM/TRIGGER I/O
+
9-16V
DC PWR
ALARM/TRIGGER I/O
Connector.
MA (Master Alarm)
output is logic low when
any channel is in alarm;
TO (Trigger Output)
output is logic low for
nominal 125 s at the
start of any scan; TI
(Trigger Input) input
logic low triggers
scanning; DC PWR (dc
volts input) input is 9 to
16V dc to power the
instrument.
1-6
Universal Input Module.
Directly wires 20 analog inputs
(Channels 1 to 20) without need
for external signal conditioning.
AC Power Connector.
Connects to any line
source of 107 to 264 volts
ac (50/60 Hz).
OVERVOLTAGE CATEGORY II PER IEC 1010-1
WARNING
ON / OFF
DIGITAL I/O
MA TO TI
0 1 2
3 4 5 6 7
+30V
MEETS VFG 243 / 1991
DIGITAL I/O
Connector.
Alarm outputs (logic
low for a channel in
alarm) and general
purpose I/O (terminals
0 to 7); totalizer input,
and GND.
Figure 1-4. 2640A/2645A Rear Panel
Ethernet 10BaseT
Connector.
A RJ-45 connector that
interfaces the instrument with
a 10BaseT Twisted-Pair
Ethernet network. The
instrument automatically
selects the active 10Base2 or
10BaseT connector.
Serial Port.
: 2640A / 41A
2645A / 46A
MODEL
: TO AVOID ELECTRICAL SHOCK, DISCONNECT LINE CORD BEFORE REMOVING COVER
SERIAL PORT
ETHERNET
Ground Terminal.
Use for 50-ohm
termination ground
lug.
Ethernet 10Base2
Connector.
A BNC-type connector that
interfaces the instrument
with a 10Base2 coax
network. The instrument
automatically selects the
active 10Base2 or 10BaseT
connector.
XMT RCV
LK
NOT FOR CONNECTION TO
PUBLIC TELEPHONE SYSTEMS
Ethernet Indicators.
XMT (transmit) blinks
red for instrument
Ethernet transmissions;
RCV (receive) blinks
red for any Ethernet
activity on the network;
LK (link) lights amber
when the Ethernet
interface is active for
the Twisted-Pair
connection, and blinks
for a data collision for
the coaxial connection.

Hide quick links: