Page 3
LIMITED WARRANTY & LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries or to any product which, in Fluke’s opinion,...
Page 4
No. 004-000-00345-4. Declaration of the Manufacturer or Importer We hereby certify that the Fluke Model 2640A/2645A Networked Data Acquisition Unit is in compliance with BMPT Vfg 243/1991 and is RFI suppressed. The normal operation of some equipment (e.g. signal generators) may be subject to specific restrictions. Please observe the notices in the users manual.
Page 5
SAFETY TERMS IN THIS MANUAL This instrument has been designed and tested in accordance with IEC publication 1010-1, Safety Requirements for Electrical Measuring, Control and Laboratory Equipment . This Users Manual contains information, warnings and cautions. Use of this equipment in a manner not specified herein may impair the protection provided by the equipment.
Page 6
DC POWER SOURCE The instrument may also be operated from a 9V to 16V dc power source when either the rear panel ground binding post or the power cord grounding conductor is connected properly. USE THE PROPER FUSE To avoid fire hazard, for fuse replacement use only a 1/4 ampere, 250V non-time delay line fuse.
Page 8
2640A/2645A NetDAQ Users Manual Preparing for Operation ... 2-1 2-1. Introduction... 2-3 2-2. Instrument Preparation... 2-3 2-3. Unpacking and Inspecting the Instrument ... 2-5 2-4. Positioning and Rack Mounting ... 2-5 2-5. Connecting to a Power Source and Grounding ... 2-5 2-6.
Page 9
2-46. Installing NetDAQ Logger with Trumpet... 2-53 2-47. Changing from an Isolated Network to a General Network ... 2-54 2-48. Installing Trend Link for Fluke (Optional) ... 2-55 2-49. Testing and Troubleshooting... 2-55 2-50. Testing the Installation... 2-55 2-51. Troubleshooting Network Problems ... 2-58 Configuring NetDAQ Logger for Windows...
Page 10
Increasing Data Transmission and Storage Rate... 4-20 4-22. Increasing Network Speed... 4-21 4-23. Optimizing Performance for Precision ... 4-21 4-24. Using Online Help ... 4-21 Using Trend Link for Fluke... 5-1 5-1. Introduction... 5-3 5-2. Displaying a Trend Link Chart During Logging ... 5-3 5-3.
Page 11
5-13. Exporting Trend Link Data Files ... 5-9 5-14. Deleting Old Trend Link Files ... 5-11 5-15. Getting the Right Look for Your Trend Link Chart ... 5-12 5-16. Using the Trend Link Control Bar ... 5-12 5-17. Using the Trend Link Menus ... 5-15 5-18.
Page 12
D RTD Linearization... D-1 E Computed Channel Equations ... E-1 Data File Format... F-1 G Dynamic Data Exchange (DDE) ... G-1 H Ethernet Cabling... H-1 Network Considerations... I-1 Error Messages & Exception Conditions ... J-1 K Fluke Service Centers... K-1 Index...
Page 14
2640A/2645A NetDAQ Users Manual A-21. 2640A Thermocouple Specifications ... A-15 A-22. 2640A Frequency Accuracy Specifications... A-16 A-23. 2640A Frequency Sensitivity Specifications... A-16 A-24. 2645A DC Voltage Measurement General Specifications ... A-17 A-25. 2645A DC Voltage Resolution and Repeatability Specifications ... A-18 A-26.
Page 15
Figure 1-1. 2640A/2645A NetDAQ Networked Data Acquisition Units ... 1-3 1-2. 2640A/2645A Front Panel ... 1-4 1-3. Typical Front Panel Display During Scanning and Monitoring ... 1-5 1-4. 2640A/2645A Rear Panel... 1-6 2-1. Instrument Preparation ... 2-4 2-2. Connecting the Instrument to a Power Source ... 2-6 2-3.
Page 16
2640A/2645A NetDAQ Users Manual 2-28. Viewing the Instrument Ethernet Address... 2-41 2-29. Examples for Viewing the Ethernet Address... 2-42 2-30. Preparing for Network Operation ... 2-44 2-31. Interconnection Using 10Base2 (Coaxial) Wiring... 2-46 2-32. Host Computer/Instrument Direct Connection ... 2-47 2-33.
Overview Introduction Introduction 1-1. The 2640A and 2645A NetDAQ Networked Data Acquisition Units are 20-channel front ends that operate in conjunction with NetDAQ Logger for Windows (hereafter known as NetDAQ Logger) to form a data acquisition system. The instruments measure dc volts, ac volts, Ohms, temperature, frequency, and dc current.
Users Manual Instrument Features and Capabilities The following describes the front and rear panels of the instrument and its capabilities (Figures 1-2 to 1-4). NetDAQ NETWORKED DATA ACQUISITION UNIT REVIEW LAST Primary, Secondary, and Annunciator Displays. Indicators and annunciators for operating mode, configuration, display, and data measurements.
REM (Remote) Annunciator. Indicates the Host Computer and the Instrument are communicating on the network, i.e., the instrument is being operated remotely. SCAN (Scanning) Annunciator. Indicates the instrument is scanning. REM SCAN m Annunciator. Indicates the multiplier for the reading is .001 (milli).
Users Manual Ground Terminal. Connects mainframe to ground. Power Switch. Applies power to the instrument (ac or dc operation). AC Power Connector. Connects to any line source of 107 to 264 volts ac (50/60 Hz). 107-264V 50/60 Hz 15VA ON / OFF ALARM/TRIGGER I/O MA TO TI 9-16V...
Analog Channels The analog channel (1 to 20) measurement connections are made via the Universal Input Module. External signal conditioning for the analog inputs is not necessary. The host computer configures all analog channels using NetDAQ Logger. Computed Channels In addition to the 20 analog channels, the instrument provides an additional 10 computed channels (21 to 30) by processing analog channels and other computed channels.
Users Manual Alarms Two alarms, Alarm 1 and Alarm 2, can be applied to any configured channel. An alarm condition occurs when a measurement falls below a low alarm value or rises above a high alarm value. You can use alarms to trigger scanning (see “Alarm Triggering”) and to set a Digital I/O line to a logic low (see “Digital I/O”...
overflows (reaching the maximum count), the display briefly shows OL (overload) and begins counting from zero again A totalizer input from contact closures increments on the “open” portion of the switch sequence close-open. To prevent switch contact “bounce” from triggering false readings, select the Totalizer Debounce feature.
Users Manual Master Alarm Master Alarm is an instrument output line that is logic low (nominal +0.8V dc) for as long as any channel is in alarm while scanning is active. The connection uses the ALARM/TRIGGER I/O terminals MA and GND (Figure 1-4). This TTL output interfaces with external equipment such as warning lights, alarms, automatic shutdowns, and paging systems.
The Spy utility gets current readings from any eight channels whether or not scanning is active. Real-Time Trend Plotting. Trend Link for Fluke (optional) lets you view real- time or historical trends in your collected data. It compares data from multiple...
Users Manual operations refer to the instrument by BCN. NetDAQ Logger supports up to 20 instruments for operation. You cannot operate an instrument from more than one host computer at a time. Isolated Networks An isolated network consists of only NetDAQ instruments and host computers. The advantages include simplified setup, faster network operation, and freedom from general network problems.
Designate one instrument in the group as the Master and the others as Slaves. The Master controls scanning operations, including the scan intervals and method of scan triggering. You can create only one group instrument. Scanning and Logging When a scan is triggered, the instrument scans the 20 analog channels and calculates the 10 computed channels.
Instrument Connector Set The 2620A-100 is a complete set of input connectors: one Universal Input Module, one ALARM/TRIGGER I/O connector, and one DIGITAL I/O connector. A 2620A-100 Instrument Connector Set comes with each instrument. You can wire additional connector sets to allow quick interfacing to multiple wiring setups.
Preparing for Operation Contents 2-1. Introduction ... 2-3 2-2. Instrument Preparation ... 2-3 2-3. Unpacking and Inspecting the Instrument... 2-5 2-4. Positioning and Rack Mounting... 2-5 2-5. Connecting to a Power Source and Grounding... 2-5 2-6. AC Power... 2-6 2-7. DC Power...
Page 34
2-46. Installing NetDAQ Logger with Trumpet ... 2-53 2-47. Changing from an Isolated Network to a General Network... 2-54 2-48. Installing Trend Link for Fluke (Optional)... 2-55 2-49. Testing and Troubleshooting ... 2-55 2-50. Testing the Installation ... 2-55 2-51.
Introduction This chapter describes how to prepare the instruments, host computers and network for operation, and how to test and troubleshoot system operation. Setting up your system requires the following steps (described in detail later in this chapter), performed in the order shown: Instrument Preparation Unpacking and setting up the NetDAQ instrument.
Users Manual Unpacking and Connecting to a Power Source Input/Output (I/O) Connections REVIEW LAST AUTO Control and Indicators Inspection SCAN FUNC Mx+B ALARM C F RO mV AC DC LIMIT HI x1Mk Figure 2-1. Instrument Preparation NetDAQ NETWORKED DATA ACQUISITION UNIT COMM ENTER NetDAQ...
You can connect the instrument to an ac power source between 107 to 264V ac (45 to 65 Hz), to a dc power source between 9 and 16V dc, or to both. Fluke guarantees equipment specifications only for 50 Hz and 60 Hz operation. Refer to Figure 2-2 and the descriptions below for making power connections.
Plug the line cord into the connector on the rear of the instrument as shown in Figure 2-2. The instrument operates on any line voltage between 107 and 264V ac (45 to 65 Hz) without adjustment. Fluke warrants the instrument to meet specifications only at 50 Hz and 60 Hz operation. Power consumption is a nominal 15 watts.
DC Power The instrument operates from any dc voltage between 9 and 16 volts. Power consumption is a nominal 6 watts. To connect the ALARM/TRIGGER I/O connector to the rear panel, complete the following procedure: 1. Remove the ALARM/TRIGGER I/O connector from the packing material or instrument rear panel.
Page 40
Users Manual 5. Close the module cover, secure the screws, and insert the module in the connector at the rear of the instrument until it latches in place. Resistance and RTD measurements use two terminals (one channel) or four terminals (two channels). The 4-wire connection provides increased accuracy over the 2-wire connection.
H L H L H L H L H L H L H L H L H L H L H L H L H L H L Figure 2-3. Universal Input Module Connections Instrument Preparation STRAIN RELIEF H L H L H L...
Users Manual SOURCE (4-WIRE) SENSE (4-WIRE) Use H and L terminals for any channel. • Channels 1 through 20 on rear panel input module (Channel 8 shown here). SOURCE (4-WIRE) SENSE (4-WIRE) Use H and L terminals for two channels on rear panel input module. Connections for Channel 8 are shown here with Channel 18 providing the additional two connections.
Shielded Wiring Use shielded wires and sensors (such as thermocouples) in environments where electrical noise is present, and connect the wire shield to the chassis ground terminal. Also refer to Appendix B "Noise, Shielding and Crosstalk Considerations." Crosstalk Considerations Crosstalk between measurement lines causes one signal to interfere with another, introducing measurement errors.
Users Manual 2. Loosen the wire clamp screw for the associated terminal. 3. Feed the wire into the gap between the connector body and the wire clamp. 4. Tighten the wire clamp; do not overtighten and crush the wire. 5. Repeat steps 2 through 4 for each wire. 6.
Preparing for Operation Instrument Preparation Complete the following procedure to make a connection to the ALARM/TRIGGER I/O connector: 1. Remove the ALARM/TRIGGER I/O connector from the rear panel. 2. Loosen the wire clamp screw for the associated terminal. 3. Feed the wire into the gap between the connector body and the wire clamp. 4.
Users Manual Instrument Connector DC Positive Input DC Negative Input Master Alarm Output Trigger Out Output Trigger In Input Signal Ground Trigger Output Trigger Output uses terminals TO and GND, and is a TTL signal that goes to a logic low for 125 µs every time a scan begins. Use the Trigger Output to trigger other instruments via their Trigger Input connection and to interface with external equipment.
External Trigger Wiring for a Group Instrument External Trigger Wiring for a group instrument refers to the triggering configuration in which you connect the Master TO (Trigger Out) line to each Slave TI (Trigger In) line and provide a common connection to the GND line for each instrument.
Users Manual 107-264V 50/60 Hz 15VA ALARM/TRIGGER I/O MA TO TI 9-16V DC PWR External trigger if used 107-264V 50/60 Hz 15VA ALARM/TRIGGER I/O MA TO TI 9-16V DC PWR 107-264V 50/60 Hz 15VA ALARM/TRIGGER I/O MA TO TI 9-16V DC PWR Figure 2-7.
Front Panel Controls Use the front panel controls (Figure 2-8) to enter configuration parameters, and choose monitoring functions. Table 2-1 summarizes the front panel control functions. Display Digital I/O and Totalizer Status Set/Review COMMunication Parameters Table 2-1. Front Panel Key Descriptions COMM Communication - Set up communication parameters.
Users Manual Front Panel Indicators The front panel indicators (Figure 2-9) consist of two five-digit displays and a set of annunciators. Table 2-2 summarizes the front panel indicator functions. REVIEW LAST Primary Display Table 2-2. Annunciator Display Descriptions Annunciator REVIEW Displays while reviewing the instrument parameters.
Page 51
Table 2-2. Annunciator Display Descriptions (cont) Annunciator °C Displays when you monitor a channel for which the measurement function is in degrees Celsius. °F Displays when you monitor a channel for which the measurement function is in degrees Fahrenheit. (Not Used.) Displays when you monitor a channel for which the measurement value is scaled by .001 (milli).
Users Manual Rear Panel Controls The rear panel has a single control: the power switch (Figure 2-10). The power switch controls both ac and dc power inputs. Rear Panel Indicators The rear panel has three LED indicators for the Ethernet adapter (Figure 2-11). Red LED blinks for instrument transmitting Ethernet data.
Front Panel Operating Procedures Power-On Options There are three power-on options as listed below: Normal Power-On Turn power switch on. The instrument communication parameters are the same as when the instrument was last turned off. Configuration-Reset Power-On Hold the front panel COMM key down, and then turn the power switch on.
Users Manual Displaying a Monitor Channel Perform the procedure in Figure 2-12 to monitor an instrument analog channel (01 to 20) or computed channel (21 to 30). See Figure 2-13 for examples. Channel Display When you press the MON key, the first monitor channel displayed is the channel most recently monitored.
Monitor display for 13.758 mV DC, GCN (Global Channel Number) 511 Monitor display for Scale Overload V AC (reading is greater than the selected range), GCN 4507 Monitor display for 234.96 F (Thermocouple), GCN 512 (otc displays for open thermocouple) Monitor display for 23.884 FUNC (Computed Channel), GCN 522 Monitor display for analog channel 18 with Mx+B scaling, GCN 818 Figure 2-13.
Users Manual Displaying the Digital I/O Status Perform the procedure in Figure 2-14 to display an instrument Digital I/O line status. The instrument updates the DIO display once per second. (See Figure 2-15 for examples.) Number Of DIO Lines There are eight DIO lines: DIO 0 to DIO 7. You can assign DIO lines as alarm outputs or as digital inputs.
Preparing for Operation Instrument Preparation Digital I/O status display for DIO line 7 (for the example 1111-0000) Digital I/O status display for DIO line 4 (for the example 1111-0000) Digital I/O status display for DIO line 0 (for the example 1111-0000) Totalizer status display for the high digits (for the example 4294967295) Totalizer status display for the low digits (for the example 4294967295) Figure 2-15.
Users Manual Displaying the Totalizer Status Perform the procedure in Figure 2-16 to display the instrument Totalizer status. The instrument updates the Totalizer display once per second. (See Figure 2-15 for examples.) To clear the Totalizer count, cycle the instrument power. You can also configure NetDAQ Logger to clear the Totalizer count when it starts logging.
Reviewing and Setting the Base Channel Number Perform the procedure in Figure 2-17 to review or set the Base Channel Number (BCN). The BCN identifies the instrument. The BCN is also the first two digits of the Global Channel Number (GCN), which uniquely identifies each instrument channel.
Users Manual REVIEW Communications display for reviewing the Base Channel Number (BCN) Base Channel Number display for setting the BCN 10s digits (for example, 45) REVIEW Base Channel Number display for reviewing the BCN number (for example, 45) Front Panel display for an instrument with BCN 45 Figure 2-18.
Reviewing and Setting the Line Frequency Perform the procedure in Figure 2-19 to review or set the line frequency. Line frequency selection allows the instrument to optimize internal circuitry for maximum precision. (See Figure 2-20 for examples.) Line Frequency Choices Select 50 Hz or 60 Hz as the frequency of the primary power when an ac source powers the instrument.
Users Manual REVIEW Communications display for reviewing the line frequency Communications display for setting the line frequency Line frequency display for setting the line frequency to 60 Hz Line frequency display for setting the line frequency to 50 Hz REVIEW Line frequency display for reviewing the line frequency (60 Hz) Figure 2-20.
Reviewing and Setting the Network Type Perform the procedure in Figure 2-21 to review or set the network type to isolated. Perform the procedure in Figure 2-23 to review or set the network type to general. An isolated network consists of only NetDAQ instruments and one or more host computers.
Users Manual REVIEW Communications display for reviewing the network type Communications display for setting the network type Network display for setting the network type to isolated Network display for setting the network type to general REVIEW Network display for reviewing the network type (isolated network) Figure 2-22.
If you install NetDAQ Logger for general network operation, you must set the network type of each instrument to general. You will need to enter an IP address, socket port, and possibly a subnet mask and gateway address into each instrument. Get this information from your network administrator.
Users Manual REVIEW Communications display for reviewing the network type Communications display for setting the network type Network display for setting the network type to general Socket Port display for setting the first digit (for the example 04369) Socket Port display for setting the second digit (for the example 04369) Figure 2-24.
Page 67
IP address display for setting an IP:0 digit (for example, 129:196:152:101) IP address display for setting an IP:1 digit (for example, 129:196:152:101) IP address display for setting an IP:1 digit (for example, 129:196:152:101) IP address display for setting an IP:2 digit (for example, 129:196:152:101) IP address display for setting an IP:3 digit (for example, 129:196:152:101) Figure 2-24.
Users Manual Reviewing and Setting the General Network Socket Port Perform the procedure in Figure 2-25 to review or set the general network Socket Port (1024 to 65535). The default is 04369. In order to communicate with each other, a host computer and an instrument must use the same socket port number. (See Figure 2-25 for examples.) General Network Socket Port Enter the Socket Port supplied by your network administrator.
Reviewing and Setting the General Network IP Address Perform the procedure in Figure 2-26 to review or set the instrument’s general network Internet Protocol (IP) address. (See Figure 2-24 for examples.) General Network IP Address Enter the IP Address supplied by your network administrator and recorded inside the rear cover of this manual for each BCN.
Users Manual Reviewing and Setting the Subnet Mask and Default Gateway If communication between the host computer and the NetDAQ instrument passes through a router or gateway, you must set the subnet mask and default gateway address on both the host computer and the instrument. Get this information from your network administrator.
Page 71
Preparing for Operation Instrument Preparation Subnet Mask The subnet mask is a 32-bit binary number expressed as four 3-digit segments, like an IP address. The subnet mask, when masked with the instrument IP address, determines what the network number is. For example, if the IP address is 129.196.180.93 and the subnet mask is 255.255.255.0, the network number is 129.196.180.0.
Users Manual Press the COMM key to review the parameters, or press and hold the COMM key COMM for 3 seconds to set the parameters. Press the up/down arrow keys until dgAtE (default gateway) appears in the primary display (COMM appears in the secondary display). Press the ENTER key.
Ethernet Address Format The Ethernet address is a 12-digit hexadecimal number. For example, 00:80:40:12:34:56. The first 6 hexadecimal digits represent a manufacturer, for example, 00:80:40 represents Fluke Corporation. The remaining digits are a sequential number assigned during manufacturing. Ethernet addresses are always unique; they are never altered, reused, or duplicated.
Users Manual REVIEW Communications display for viewing the instrument Ethernet address REVIEW Ethernet address display for viewing byte 0 (for the example 00-80-40-12-34-56) REVIEW Ethernet address display for viewing byte 2 (for the example 00-80-40-12-34-56) REVIEW Ethernet address display for viewing byte 4 (for the example 00-80-40-12-34-56) REVIEW Ethernet address display for viewing byte 5 (for the example 00-80-40-12-34-56) Figure 2-29.
Host Computer and Network Preparation This section contains information for preparing your host computer and setting up network communication, as summarized in Figure 2-30. Installing Host Computer Ethernet Adapter Skip this section if you have an Ethernet adapter installed on your computer. Since the installation procedures for Ethernet adapters change frequently and without notice, you must follow the instructions supplied with your particular Ethernet adapter.
Install Logging Software Install Trending Software (Optional) Figure 2-30. Preparing for Network Operation 2-44 PCMCIA Card Parallel-to-LAN Adapter NetDAQ NETWORKED DATA ACQUISITION UNIT COMM ENTER Instrument Windows 95/NT Trumpet NetManage Newt Fluke NetDAQ Logger for Windows Trend Link for Fluke...
Instrument and Host Computer Interconnection You may interconnect NetDAQ instruments and host computer(s) with either 10Base2 (coaxial) or 10BaseT (twisted pair) wiring. If your site is already wired, you will probably use the wire in place. If your site is not wired, and you are connecting your instrument directly to your host computer, it is easiest to use the coaxial cable supplied with your NetDAQ instrument.
Users Manual Interconnection Using 10Base2 (Coaxial) Wiring Connect the ground lug on the 50-Ohm terminator to the instrument ground terminal only when there is no other shield ground connected for the network. 10Base2 interconnection uses 50-Ohm coaxial cables (Belden 9907 or equal) that loop between equipment items and connect at each station through a BNC "T"...
Users Manual Installing Host Computer Networking Software To establish Ethernet communication in your host computer, you must do the following: Install a driver for the adapter Install a TCP/IP protocol stack and Windows Socket (Winsock) software Set host computer networking parameters This section discusses installing the adapter driver and the TCP/IP protocol stack and Winsock software.
Preparing for Operation Host Computer and Network Preparation Setting Host Computer Networking Parameters 2-44. This section discusses how to set your host computer networking parameters after you install your adapter and networking software. I f you plan to install NetDAQ Logger for general network operation, and you are just now enabling networking, you must set the host computer’s IP address, subnet mask, and possibly its default gateway IP address.
4. Select the language of the user interface: English, French, German, or Spanish. Click Next. 5. Select the program folder. The default name of the program folder or program group is Fluke NetDAQ Logger. You can select or type a different name. Click Next. 2-52 2-45.
6. Check the setup. If the setup is correct, click Next to begin file transfer. If you want to change the setup, click Back to go to the appropriate screen and make the changes. Click Next and check the setup again. If it is correct, click Next to begin file transfer.
Click Next. 7. Select the program folder. The default name of the program folder or program group is Fluke NetDAQ Logger. You can select or type a different name. Click Next. 8. Check the setup. If the setup is correct, click Next to begin file transfer.
Computer Networking Parameters” to configure the host computer for general network operation. Installing Trend Link for Fluke (Optional) The Trend Link setup program automatically determines whether to install the 32-bit (for Windows 95 and Windows NT) or 16-bit version of the software. To install Trend Link, complete the following procedure: 1.
Page 88
Users Manual Setting line frequency and network parameters on the instrument if the defaults are not applicable Installing an Ethernet adapter in your host computer Interconnecting the host computer and the instrument Installing TCP/IP software Installing NetDAQ Logger Installing Trend Link (optional) Complete the following procedure: 1.
Page 89
2. Select Setup | Communications Config. 3. Click Add to open the Instruments on Network dialog box to add a NetDAQ instrument to NetDAQ Logger’s list of instruments. Select the instrument model number and enter the instrument BCN. On a general network, you must enter the instrument IP address which you previously recorded inside the back cover of this manual.
Users Manual Troubleshooting Network Problems Review the troubleshooting information below to help locate any network problems. Table 2-4 summarizes network messages reported by NetDAQ Logger for Windows. Table 2-5 is a summary of how to use the Ethernet LED indicators on the instrument to identify network problems.
Page 91
Table 2-4. Network Error Messages (cont) Reported Message Connection is down! The host computer cannot establish network communication. Unable to initialize The host computer Winsock! software cannot find or open the Winsock DLL. Comment 2. Try Verify Communications. If it still fails, then either the host computer Ethernet adapter is not working or the instrument is defective.
Problem The Ethernet Parallel-to-LAN Adapter is not responding -or- Mouse or Keyboard is not responding, or software behaves abnormally. Unable to verify communications with an instrument. General Protection Faults. Changed instruments on the network and the new instrument do not respond. Table 2-6.
Introduction NetDAQ Logger provides an easy method for building configuration databases, transferring the configuration to the instrument, collecting data from the instrument, and managing the collected data. This chapter provides instructions on managing configuration information for NetDAQ Logger. Chapter 4 provides details on operating NetDAQ Logger: starting and stopping the instruments, managing data files, and optimizing performance.
Users Manual Configuring Network Communications Before you can configure the instruments on your network and create setup files for these instrument configurations, you must configure your network communications. This involves identifying the instruments by their BCN numbers and, on a general network, their IP addresses. Network configuration data is not saved as part of the setup files but is its own separate file, ccf.cfg, maintained by NetDAQ Logger.
The Communications Configuration Dialog Box Use the Communications Configuration dialog box (Figure 3-2) to record your network configuration. To open the Communications Configuration dialog box, select Setup | Communications Config or click the Communications Configuration toolbar button. The dialog box lets you view the instruments currently on the network, modify them, and add or delete instruments.
Page 100
Users Manual 3. Click OK to return to the Communications Configuration File dialog box, which now lists the new instrument. For a general network, the IP address appears after the instrument model number. 4. If you are on a general network, you may need to change the socket port if a conflict occurs with other network operations.
Deleting an Instrument from the Network Complete the following procedure to remove an instrument from the Instruments on Network List, which is common to all setup files. Any instruments you delete in this procedure will no longer be accessible to any setup file. Do not use this procedure simply to remove an instrument icon from the Icon Bar.
Users Manual Configuring the Current Setup NetDAQ Logger keeps track of all instrument configuration information (e.g., reading rate and measurement functions) for all the instruments represented on the icon bar. This collection of information is called the current setup. NetDAQ Logger allows you to save the current setup to a file and open previously saved setup files.
Deleting an Instrument Icon To delete an instrument icon, select Setup | Delete Instrument Icon. A message appears warning you that the instrument’s configuration will be lost. If you choose to delete the icon, it will disappear from the Icon Bar in the Main Window. Designating Instruments as Group or Asynchronous You can designate instruments as grouped or asynchronous (standalone).
Page 104
Users Manual When a grouped instrument is externally wired, you can select any combination of trigger types for the Master instrument: Interval Trigger, External Trigger, or Alarm Trigger. NetDAQ Logger configures the Slaves for External Trigger so they will respond to the trigger signal from the Master. Unwired Group If you cannot wire the grouped instrument, select only the Interval Trigger for the Master.
Setup Files NetDAQ Logger allows you to save the current setup in a file. You can open the setup file later to reestablish the saved setup. Using setup files, you can create any number of setups in advance and use them when applicable. You can set up your system so that you can activate NetDAQ Logger by opening a setup file.
Users Manual Opening a Setup File If NetDAQ Logger is already active, you can load a saved setup into the current setup by opening a setup file. Before you open a setup file, you must stop logging on all instruments. Use any of the following methods to open a setup file. Select Setup | Open Setup to open the Open Setup dialog box.
Starting NetDAQ Logger with a Setup File On Windows 95, there are several ways to start NetDAQ Logger and open a specific setup file at the same time: In the Windows Explorer or My Computer, you can double-click on the name of the desired setup file.
Users Manual 3. On the Shortcut tab, add the setup file name to the end of the command line in the Target box (see “NetDAQ Logger Command Line” for the command lines). You can use Start | Settings | Taskbar to modify the command line of NetDAQ Logger to open a setup file whenever you start NetDAQ Logger from Start | Programs.
To activate NetDAQ Logger with the default setup: “pathname\netdaq32” To activate NetDAQ Logger with a specific setup file (for example, mysetup.stp): “pathname\netdaq32” “mysetup.stp” To activate NetDAQ Logger with a specific setup file and start logging automatically: “pathname\netdaq32” “mysetup.stp” /g To add calibration to the Utilities menu of NetDAQ Logger: “pathname\netdaq32”...
Users Manual Figure 3-3. Instrument Configuration Dialog Box 1. Select Celsius or Fahrenheit to use as temperature units for thermocouple and RTD temperature measurements. 2. Select a monitor channel. This channel will automatically be displayed on the instrument’s front panel when logging is started. 3.
Page 111
External Only A scan is triggered immediately after the external trigger input is pulled low (active). While the external trigger line remains low, each successive scan is triggered after Interval 2 elapses since the start of the previous scan. Alarm Only A scan is triggered when the instrument detects an alarm condition on a channel being used as an alarm trigger.
Users Manual Configuring Channels These procedures provide instructions for configuring an instrument’s analog channels and computed channels. These instructions include selecting the measurement function, enabling Mx+B scaling, configuring alarms, and entering channel labels. The Channels Configuration Dialog Box All channels configuration procedures start with the Channels Configuration dialog box shown in Figure 3-4.
Configuring Analog Channel Functions Analog channel functions include volts dc, volts ac, ohms, temperature (thermocouple and RTD), frequency, and dc current. To configure analog channel functions, complete the following procedure: 1. Select an analog channel in the Channels Configuration dialog box. 2.
Users Manual Configuring Computed Channel Functions Computed channel functions include average, difference, difference from average, and equation. To select computed channel functions, complete the following procedure: 1. Select a computed channel in the Channels Configuration dialog box. 2. Click Function to open the Computed Function dialog box. 3.
3. If the equation contains an error, an error message appears. Click OK in the error message box. If possible, NetDAQ Logger highlights the location of the error in the equation. Change the equation as necessary and click OK to return to the Channels Configuration dialog box.
Users Manual Allowed (-2*c21/c2)/(2*((c1+1)-(c21/c2)*(c1-1))) See Appendix E for syntax definition and more detailed information. Configuring Mx+B Scaling Mx+B scaling multiplies a measurement by a multiplier (M) and then applies an offset (B). For example, Mx+B scaling of 100x+50 applied to a measured value of 1.15 would result in a reading of 100(1.15)+50=165.
Configuring Alarms Each channel has two alarms, each of which may be set to HI, LO, or OFF. An alarm condition occurs when a measurement falls below a low alarm value (LO alarm) or rises above a high alarm value (HI alarm). NetDAQ Logger records all alarm conditions in the data file.
Users Manual 3. Select a digital output if you would like to associate an alarm with a digital I/O line. The instrument will set the digital I/O line to a TTL logic low for as long as the alarm conditions is met. 4.
The values in the Mx+B file must have the following format: <channel number> <M value> <B value> <cr> <lf> Each parameter is separated by spaces or tabs and each definition ends with a carriage return (<cr>) and a line feed (<lf>). The decimal point character corresponds to your Windows International settings.
Users Manual Copying a Channels Configuration Complete the following procedure to copy a complete instrument channels configuration from one instrument to another instrument. This includes channel Functions, Mx+B Scaling, Alarms, and Channel Labels. For example, if you set up three instruments with identical channel configurations, you can configure one instrument and then copy the channels configuration to the other two instruments.
Default Configuration Settings When you create a new instrument icon, NetDAQ Logger initializes the settings for that instrument to the default parameters. (See Table 3-1.) Table 3-1. NetDAQ Logger Default Instrument Configuration Parameter Description: Group Status: Trigger Type Interval 1: Interval 2: Reading Rate: Drift Correction:...
Users Manual Using Configuration Lockout Use The Configuration Lockout feature to prevent accidental instrument configuration changes. The following commands in the Setup menu will be disabled: Create Instrument Icon, Delete Instrument Icon, Group Instruments, Communications Config, Instrument Description, Instrument Configuration, Load MxB+ File, and Copy Channels.
NetDAQ Logger to put the changes into effect. Refer to the default netdaq.ini file shown below. [Defaults] MemoryWrapAround=True AlarmCheckingIntvl_2640A=1.000 AlarmCheckingIntvl_2645A=0.100 AutoScanDisable=True CommTimeout=10 [TL Defaults] TLDir=c:\Program Files\Fluke\tl TLFile=tl32.exe TLTitle=Trend Link for Fluke Configuring the netdaq.ini File (None) 2640A Interval 1.000 sec Slow (High Res.) Celsius None Alarm 1 Alarm 2...
Page 124
Plot/Trend | Trend Link Directory. TLFile=tl32.exe This is the file in TLDir that executes when the you select Plot/Trend | Show Trend Link. TLTitle=Trend Link for Fluke This is the title of the Trend window that appears when TLFile executes. 3-30...
Operating NetDAQ Logger for Contents 4-1. Introduction ... 4-3 4-2. Starting and Stopping Logging... 4-3 4-3. Starting or Stopping all Instruments at Once ... 4-4 4-4. Starting or Stopping a Group Instrument ... 4-4 4-5. Clearing an Instrument’s Totalizer Value... 4-5 4-6.
Introduction This chapter provides instructions for operating your system using NetDAQ Logger after you have set up communications and configured the instruments. This includes starting and stopping logging, viewing and storing your data, and optimizing system performance. Starting and Stopping Logging You can start or stop logging for an individual instrument, a group instrument (by starting or stopping the Master), or all instruments.
Users Manual To stop logging for an individual instrument, complete the following procedure: 1. Select the instrument’s icon. 2. Select Logging | Stop Instrument or click the Stop Instrument button on the toolbar. If the menu command and button are dimmed, the instrument is not logging, or it is a Slave and you must stop the Master.
Clearing an Instrument’s Totalizer Value If you check Logging | Clear Totalizer on Start, NetDAQ Logger will clear the totalizer count in each instrument when you start logging for that instrument. Otherwise, the totalizer value in the instrument will not be cleared. Simulated Logging If you check Logging | Simulate Logging on Start, NetDAQ Logger will simulate all data collection rather than connecting to the instrument(s).
Users Manual 3. Check Stop Scanning after Interval. Enter the hours/minutes/seconds for the duration. The maximum entry is 999 hours, 99 minutes, 99 seconds (999:99:99). 4. Click OK to return to the Main Window. Real-Time Displays The Real-Time displays let you monitor data as it is being collected. The following displays are available: Logging Status displays the status of data collection during real scanning or simulated scanning.
Trend Link provides a wide range of features for trending real time data or previously collected data. See Chapter 5 of this manual for more information on Trend Link. Dynamic Data Exchange (DDE) links measurement data to other applications such as a spreadsheet, with the measurements appearing in other applications as they are being collected.
Users Manual The Readings Table Window The Readings Table provides a numerical display of the most recent data collected from the currently selected instrument. Complete the following procedure to use the Readings Table window: 1. Select Options | Show Current Readings to display the Readings Table window.
2. Select the icon of the instrument for which you would like to view incoming data. If this instrument is not currently logging, select Logging | Start Instrument. The Spy Window The Spy Window displays up-to-date measurements whether or not instruments are scanning.
Page 134
2640A/2645A NetDAQ Users Manual ds059s.bmp 2. Click OK to display the Spy window. Double-click in the Spy window to return to the Spy Channel Selection dialog box. ds060s.eps 4-10...
Quick Plot Quick Plot provides a quick way to get a simple graphical view of your logging data. Quick Plot samples the incoming scan data at a specified interval and displays a graphical trend of the data. For a more detailed view of your data, use the Trend Link application.
Users Manual Y-Axis Plot Minimum Plot Maximum X-Axis 40 data points nn seconds apart where nn is the Sampling Interval (1 to 3600 seconds) Dynamic Data Exchange (DDE) Operations Complete the following procedure to set up typical Dynamic Data Exchange operations.
Items: ‘Cnn’ (nn is the channel number 01 to 30) DIO (Digital I/O status, decimal equivalent of the 8 I/O lines) Totalizer (Totalizer count, 0 to 4294967295) Alarm1 (Alarm 1 status, decimal equivalent of the 30 Alarm 1 states) Alarm2 (Alarm 2 status, decimal equivalent of the 30 Alarm 2 states) ELDateTime (Scan date/time in Excel/Lotus 1, 2, 3 format) Begin data collection, observing the data display in the client application.
Users Manual Trend Link The Trend Link data file format requires using the real-time trending capability offered by the Trend Link trending package. You can record data into Fast Binary and then convert it into the Trend Link format. When using the Trend Link format, you may specify a deadband value for each channel.
Page 139
3. Select the desired File Format. For the Fast Binary or ASCII (CSV) format, select Append to File or Overwrite File. This file mode determines what happens when logging starts if a data file of the same name exists. Enter text characters (79 maximum) into the File Comment box.
Page 140
Users Manual 6. Select the desired Auto Rollover type and its associated value. The data file may be automatically rolled over when it reaches a specified size, when a time interval elapses from the last rollover, or at a particular time each day. To safeguard your data when recording in the Trend Link format, you must perform a rollover or stop logging data before your disk completely fills up.
Page 141
12. Enter a batch name and the desired batch definition and click OK to return to the Data File Configuration dialog box. Use the batch definition to tell NetDAQ Logger when to mark the start and stop points of batches: Entire logging session.
Users Manual Converting Data Files You can convert data files to other types for compatibility with your data management applications. NetDAQ Logger performs the following conversions: Binary to ASCII ASCII to Text (timestamp conversion) Binary to Trend Link Complete the following procedure to perform one of these conversions. This conversion does not modify or destroy the original file, but creates an additional file in the new format.
Page 143
The example below is a typical data file in text format. Off-scale measurements are recorded as +1.0E+9 (+overload) and -1.0E+9 (-overload). Open thermocouple measurements are recorded as +9.0E+9. Data File Name Measurement Scan Records Number of Columns (An asynchronous instrument has a single digit for the Number of Columns.
Users Manual Optimizing Performance You can set up instrument operations to optimize either your system’s data collection rate or measurement precision, depending on your requirements. The procedures in this section describe the options for optimizing performance. Optimizing Performance for Speed To optimize the data collection rate of your system, you can alter the parameters that affect the instrument’s scanning rate, the data put through from the instrument to the data file, or the speed of the network.
In addition, try to avoid running other tasks on the computer when you are writing to the data file. The following suggestions are: Avoid using the Quick Plot, Spy, Readings Table, or Logging Status windows. Avoid using the Trend Link window in real time. Close any other applications that are open.
Using Trend Link for Fluke Contents 5-1. Introduction ... 5-3 5-2. Displaying a Trend Link Chart During Logging... 5-3 5-3. Playing Back a Trend Link File in Trend Link... 5-4 5-4. Playing Back a Fast Binary File in Trend Link... 5-5 5-5.
Introduction Trend Link for Fluke (Trend Link) graphs real time and historical data in the Microsoft Windows environment. This chapter supports Trend Link operation with a series of procedures to use in conjunction with NetDAQ Logger. Refer to the Trend Link for Fluke Reference Manual for complete information. Make sure that you have installed Trend Link as described in Chapter 2, "Installing Trend...
Users Manual If this is the first use of the Trend Link data file, allow a few seconds for Trend Link to create the folder and files used for the chart. 4. To change the look of the chart, see "Getting the Right Look for your Trend Link Chart"...
3. Select the chart file to play back. For example, select c:\Program Files\Fluke\netdaq\_data.set\data.cht 4. Click OK. 5. Observe that the selected chart opens. Use the scroll bar at the bottom of the chart to display the measurement area of interest.
You can import standard *.csv, Iconics Format (*.prn), and Fluke Format *.csv data files using the Trend Link Import *.CSV File utility. For example purposes, the procedure below shows how to import a Fluke Format ASCII (CSV) data file. You can adapt this procedure to standard *.csv and Iconics Format *.prn as required.
Target Data Set box. For example, double-click data.set. Import File Type Select Fluke Format (*.csv) Filename Enter the complete path to the NetDAQ ASCII (CSV) file you want to play back in Trend Link. For example, enter c:\Program Files\Fluke\netdaq\testdata.csv.
Users Manual Examine the chart and perform any standard Trend Link procedures to view, print, and so forth. (See "Getting the Right Look for Your Trend Link Chart.") Title the Trend Link Chart To open the Chart Title dialog box, select Preferences | Chart Title. Enter the desired title and click OK.
To save the Trend Link chart, select File | Save As. Enter the path to your Data Set directory for the *.cht chart. For example, enter c:\Program Files\Fluke\netdaq\_data.set\data.cht. Click OK. Exporting Trend Link Data Files Trend Link can export data files into one of three Export File Type formats: Standard Comma Separated Values (*.csv), MathCad Array Format (*.prn)
Page 156
Export File Type, Sample Rate time resolution of the exported data file in the format Hours:Minutes:Seconds, and the name of the exported file. The example below shows the data.set file exported in Column Heading Format as file: c:\Program Files\Fluke\netdaq\ data.set\data.csv with a sampling interval of 0.5 second. Click OK. 5-10 ds128s.bmp...
Also, delete the *.set file. Then, open the Configurator application and complete the following steps: 1. Double-click the c:\Program Files\Fluke\tl\infolink.ini file to display the contents using the Notepad accessory. Delete the line under the heading [TL Data Set Paths] that relates to the directory you deleted above.
Using the Trend Link Control Bar The Control Bar in the Trend Link window (below) allows you to quickly modify the chart display. See the Trend Link for Fluke Reference Manual for more information on menu items and other Trend Link capabilities. 5-12 5-15.
Page 159
Using Trend Link for Fluke Getting the Right Look for Your Trend Link Chart Control Bar Historical Mode: Click to freeze the chart during the Real Time Mode, allowing you to use the scroll bar at the bottom of the chart to back up to any portion on the recorded chart.
Page 160
Users Manual Erasing a Curve To erase a curve, click this button, hold the mouse button down, drag the Eraser icon on top of the Curve icon, and release the mouse button. Adding Notes To enable the Note icon on the Control Bar, select Preferences | Note System.
This section does not discuss menu functions repeated on the Control Bar. (See "Using the Trend Link Control Bar.") This section only summarizes the menu functions. Refer to the Trend Link for Fluke Reference Manual for complete menu information.
To change the default value so that all created charts will have a selected preset value in the Time Interval Window, use the Notepad accessory to modify c:\Program Files\Fluke\netdaq\tldef.cht. Find [Section: Curve Window], and add the line Curve Window Interval Time=nnnnnnn, where nnnnnnn is the time in milliseconds.
Configuring the Curve Status Display Complete the following procedure to select the elements of the Curve Status display. The Curve Status display is the window at the bottom of the chart containing the parameter information of the curves. Toggle Curve Status on and off with the Curve Status Window button on the Control Bar.
Users Manual 2. Select the desired features of the Curve Status display by checking or unchecking each of the Sections Displayed listings. Click OK. Curve Preferences Complete the following procedure to select the curve preferences. You can customize various attributes of each curve on the chart. For example, you can change the color, add a fixed range, and position one curve on top of another.
Page 165
Using Trend Link for Fluke Getting the Right Look for Your Trend Link Chart ds135s.bmp 2. Click on the desired curve listed in the Tags box, which will then appear in the Tagname box. Click OK to open the Curve Parameters box (below). (You can also open this dialog box directly by double-clicking the Curve icon.)
Page 166
Users Manual 3. The tag name of the selected curve appears in the dialog box caption. Each dialog box entry has the following characteristics: Description This entry will be initialized with the string entered in NetDAQ Logger for the channel label. (See "Assigning Channel Labels" in Chapter 3.) Engineering Units This entry will be initialized with the string entered in NetDAQ Logger for the units label.
the left side of the curve window will include a display of scale values. Otherwise, a band without numbers will appear. Curve Type Select the type of curve: normal, highlight, shaded, or SPC (Statistical Processing Control). If you enabled the NetDAQ Logger channel alarm values, the default curve type is "highlighted"...
Page 168
Users Manual 2. The dialog box entries are as follows: Bkgrnd Color: Choose the background color. Grid Color: Choose the grid color. Display Grid: This is the same as selecting Grid Lines On/Off on the Control Bar. Calculate Interval:Select this option to enter the number of vertical grid lines to display.
Constant Interval: Select this option to enter the exact time interval that a vertical grid line should represent. Pixel Width: Select this option to enter the distance between vertical grid lines in pixels. Use scale tickmarks: If you select this option, horizontal grid lines draw to match the tickmark labels on the band bars.
Users Manual 1. Select Preferences | Chart Title to open the Chart Title dialog box. Enter the chart title name in the text box, for example, station one. Click OK. 2. Observe that the entered text appears at the top of the chart in the Title Bar. Using the Note System Complete the following procedure to add a text note to a curve.
Page 171
Using Trend Link for Fluke Getting the Right Look for Your Trend Link Chart ds140s.bmp You can position and size the printed chart by entering X and Y Offsets and Width and Height settings. The X and Y Offsets specify the position of the upper- left corner of the printed page, and the Width and Height determine the size of the printed chart.
Page 172
2640A/2645A NetDAQ Users Manual ds141f.bmp Click OK to print the chart. 5-26...
Service Manual, PN 942615. Otherwise, package the instrument securely (using the original container, if available), and mail it to the nearest Fluke Service Center. Include a description of the problem. Fluke assumes no responsibility for damage in transit.
Users Manual Cleaning Keep the instrument dry to avoid electrical shock to personnel or damage to the instrument. To prevent damage, never apply solvents to the instrument housing. Wipe the instrument with a cloth lightly dampened with water or mild detergent. Do not use aromatic hydrocarbons, chlorinated solvents, or methanol-based fluids.
If the instrument fails a performance test, the instrument requires service or repair. To perform these tests, you will need a Fluke 5700A Multifunction Calibrator and several other pieces of equipment meeting the minimum specifications given in Table 6-2.
Users Manual Host Computer INPUT MODULE Ethernet Coaxial Cable (50-ohm) Minimum cable length is 20 inches (0.5 m). BNC “T” 50-ohm Terminator Terminator Ground Wire Figure 6-2. Performance Test Setup 13 14 15 16 17 18 19 20 SOURCE H L H L H L H L H L H L H L H L H L (4-WIRE) SENSE H L H L H L H L H L H L H L H L H L...
Initializing the Performance Test Setup Complete the following procedure to initialize the performance test setup. It is assumed you have configured the host computer and instrument as described in "Configuring the Performance Test Setup" (above). Testing begins with the instrument and host computer unpowered. This assures that at power-up self-tests are completed successfully, the correct host computer Ethernet port is activated, the host computer configuration is accurately reflected, and other background operations are completed.
Page 182
Users Manual 6. Verify Communications With the Communications Configuration File dialog box still open, select instrument 01 on the Instruments on Network list and click the Verify button. The message Connection Successful! is returned for successful communications between the instrument and host computer. If you receive an error message, refer to "Error and Status Messages"...
Accuracy Performance Tests This accuracy performance test assumes you have completed "Initializing the Performance Test Setup" above. Do not begin this test until the instrument has been temperature stabilized for a minimum of 30 minutes. Do not use the instrument front panel monitor function for performance testing; use the higher resolution Spy window at the host computer, as specified in procedures.
Users Manual 3. Verify Accuracy Configure the 5700A for the output values below and verify the Spy window measurement is between the minimum and maximum values. Change the channel 1 range as required (see Step 1). Volts DC Range 90 mV Short Circuit (Zero) 90 mV 90 mV...
3. Verify Accuracy Configure the 5700A for the output values below and verify the Spy window measurement is between the minimum and maximum values. Change the channel 1 range as required (see Step 1). Volts DC Range 5700A Output 90 mV Short Circuit (Zero) 90 mV +90 mV...
Users Manual 3. Verify Accuracy Configure the 5700A for the output values below and verify the Spy window measurement is between the minimum and maximum values. Change the channel 1 range as required (see Step 1). Volts AC Range 300 mV 300 mV 300 mV 300 mV...
Analog Channel Integrity Test Complete the following procedure to test the integrity of each analog channel (2 to 20) to verify each analog channel is capable of making measurements. 1. Configure Channel for Ohms In NetDAQ Logger, configure channels 2 (then 3, then 4, etc.
Users Manual Thermocouple Temperature Accuracy Test Ensure that the Accuracy Tests (above) have been completed before performing this test. 1. Connect a Thermocouple Remove the Universal Input Module from the instrument and connect the supplied type T thermocouple to the channel 1 terminals with the blue lead to the H terminal and red lead to the L terminal.
2-Wire Resistance Accuracy Test (2640A) Complete the following procedure to test the accuracy of the resistance function for the 2640A using 2 terminals. Measurement accuracy applies to all channels, not just the channel used for the test. (The 4-wire resistance accuracy test is more rigorous and you may wish to skip this step and continue to “4-Wire Resistance Accuracy Test.”) 1.
Users Manual Resistance Range* Short Circuit (Zero) Short Circuit (Zero) 30 k 300 k * The resistance accuracy in this table makes allowance for up to 0.1 You must add any additional lead wire resistance present in your setup to the resistance values given in this table..
Page 191
Resistance Range* Decade Resistor Short Circuit (Zero) 30 k 30 k 29 k 300 k 290 k 2.9 M * The resistance accuracy in this table makes allowance for up to 0.1 Ohm of lead wire resistance plus 0.01% decade resistance tolerance. You must add any additional lead wire resistance present in your setup to the resistance values given in this table.
Users Manual 4-Wire Resistance Accuracy Test (2640A) Ensure that the Accuracy Tests (above) have been completed before performing this test on the 2640A. 1. Connect the Resistance Source to Channels 1 and 11 Remove the Universal Input Module from the instrument and connect a cable from the Decade Resistance Source to the Universal Input Module terminals for channel 1 (Sense) and channel 11 (Source) as shown in Figure 6-4.
13 14 15 16 17 18 19 20 SOURCE H L H L H L H L H L H L H L H L H L (4-WIRE) SENSE H L H L H L H L H L H L H L H L H L (4-WIRE) : ON EX SNS...
Page 194
Users Manual 4. Verify Accuracy Configure the Decade Resistance Source for the output values below and verify the Spy window measurement is between the minimum and maximum values. Change the channel 1 range as required (see Step 2). Resistance Range Short Circuit (Zero) Short Circuit (Zero) 30 k...
4-Wire Resistance Accuracy Test (2645A) Ensure that the Accuracy Tests (above) have been completed before performing this test on the 2645A. 1. Connect the Resistance Source to Channels 1 and 11 Remove the Universal Input Module from the instrument and connect a cable from the Decade Resistance Source to the Universal Input Module terminals for channel 1 (Sense) and channel 11 (Source) as shown in Figure 6-4.
Users Manual Resistance Range Short Circuit (Zero) Short Circuit (Zero) 30 k 300 k 5. Close Spy Window. RTD Temperature Accuracy Test (Resistance) (2640A) The following RTD accuracy test applies to the 2640A and uses the 4-wire connection (see Figure 6-4). 1.
RTD Temperature Accuracy Test (Resistance) (2645A) The following RTD accuracy test applies to the 2645A and uses the 4-wire connection (see Figure 6-4). 1. Connect the Decade Resistance Source to Channels 1 and 11 Remove the Universal Input Module from the instrument and connect a cable from the Decade Resistance Source to the Universal Input Module terminals for channel 1 (Sense) and channel 11 (Source) as shown in Figure 6-4.
Users Manual 4. Verify Accuracy Insert the RTD and a mercury thermometer in a room- temperature bath. Allow 20 minutes for thermal stabilization. The value displayed on the mercury thermometer should equal the value in the Spy Window +0.25 C (2640A) or +0.80 C (2645A) plus sensor inaccuracies. 5.
8. Short Channel Inputs Remove the Universal Input Module from the instrument and apply a short to channels 1 through 8. Reinstall the Universal Input Module. 9. Start Instrument Scanning Select Logging | Start Instrument to start the instrument scanning. The instrument must be scanning to set the DIO lines. 10.
Users Manual 3. Verify Digital I/O Input for all Set Lines In sequence, individually ground each DIO line to the GND line using the DIO wires connected in Step 1. Note the change in the DIO status reported in the Spy window as follows: None grounded Reported DIO Status = 255 DIO0 grounded Reported DIO Status = 254 DIO1 grounded Reported DIO Status = 253...
2. Connect Test Leads At the DIGITAL I/O connector, connect the (Totalizer) test lead and GND test lead to a signal generator’s output terminals. Adjust the signal generator for an output of 1.5V rms sine wave at 10 Hz. 3. Open Spy Window Select the Spy command from the Utilities menu. Select 01TOTAL.
Users Manual Trigger Input Test This test checks the ability of the Trigger Input line to trigger measurement scanning. 1. Configure Trigger Input In NetDAQ Logger, configure the scan parameters for External Trigger with an Interval 2 of 1 second. Be sure Interval Trigger and Alarm Trigger are not enabled.
4. Verify Trigger Output is Enabled In NetDAQ Logger, select Setup | Instrument Config. In the Instrument Configuration dialog box, verify the Trigger Out box is checked. Click OK to return to the Main Window. (See “The Instrument Configuration Dialog Box" in Chapter 3 of this manual.) 5.
Users Manual Do not press CAL ENABLE unless you have a copy of the Service Manual and intend to calibrate the instrument. If you have activated calibration and wish to exit, press CAL ENABLE until the CAL annunciator is removed from the display (or turn the power off). Adding Calibration to the Utilities Menu To add calibration to the Utilities menu, complete the following procedure: 1.
Instrument Calibration Dialog Box The Instrument Calibration dialog box (below) sets up the RS-232 port for calibrating the instrument and provides calibration function selection. See the Service Manual (PN 942615) for the calibration procedure. RS-232 Port Click the scroll arrow to list the RS-232 port choices: COM1 to COM4.
If the instrument fails to operate, check that operating instructions presented earlier in this manual are being followed. If the problem cannot be remedied, send the instrument, postage paid, to the nearest Fluke Service Center. A list of Service 6-34 key and then apply power.
Page 207
Centers is provided in Appendix K of this manual. Be sure to pack the instrument securely; use the original container if available. Include a description of the problem. Fluke assumes no responsibility for damage in transit. For application or operation assistance or information on Fluke products, telephone: 1-800-44-FLUKE (1-800-443-5853) in USA and Canada...
Fluke offers a 1-year warranty for the 264XA-801 (Ethernet plug-in card) and 264XA-802 (Ethernet Parallel-to-LAN adapter), and 264XA-803 (PCMCIA Adapter) options. Defective units may be returned to either Fluke or to the option manufacturer for repair or replacement. The original equipment manufacturer may offer warranties beyond the warranty supplied by Fluke.
Introduction Specifications are divided into three sections. The first section contains the combined specifications that apply equally to both the 2640A and 2645A instruments. The second section contains specifications that apply only to the 2640A instrument. The third section contains specifications that apply only to the 2645A instrument.
Users Manual Table A-1. 2640A/2645A General Specifications Specification Channel Capacity I/O Lines Total Size Weight Power Safety Standards EMC Standards Serial Interface (RS-232C) Common Mode Voltage Characteristic 9.3 cm high, 21.6 cm wide, 36.2 cm deep (3.67 in high, 8.5 in wide, 14.28 in deep) Net, 4 kg (8.8 lb.) Shipping, 6.0 kg (13.2 lb.) 107 to 264V ac (no switching required), 50 and...
Page 213
Table A-1. 2640A/2645A General Specifications (cont) Specification Maximum Measurement Speed (Scanning Rates) Accuracy of Medium Scanning Rate Additional error if “Automatic drift correction” is turned off. 2640A/2645A Environmental Specifications Table A-2 provides a summary of the environmental specifications for the 2640A/2645A.
Users Manual Table A-2. 2640A/2645A Environmental Specifications Specification Warm-up Time Operating Temperature Storage Temperature Relative Humidity Altitude Vibration Shock 2640A/2645A Input/Output Capabilities The following specifications include the input/output functions, including the Digital I/O, Trigger Out, Trigger In, and Master Alarm output. Digital I/O Table A-3 provides a summary of the Digital I/O specifications for the 8 Digital I/O lines (0 to 7).
Table A-3. 2640A/2645A DIGITAL I/O Specification (cont) Specification Output Voltage - TTL Logical Zero Output Voltage - TTL Logical One Output Voltage - Non-TTL Load Zero Output Voltage - Non-TTL Load One Trigger In Table A-4 provides a summary of the Trigger In specifications. The Trigger In input is located on the ALARM/TRIGGER I/O connector, terminals TI and GND.
Users Manual Trigger Out Table A-5 provides a summary of the Trigger Out specifications. The Trigger Out output is located on the ALARM/TRIGGER I/O connector, terminals TO and GND. Table A-5. 2640A/2645A Trigger Out (TO) Specification Specification TTL Logical Zero - Trigger Out Set TTL Logical One - Trigger Out Not Set Non-TTL Logical Zero - Trigger Out Set Non-TTL Logical One - Trigger Out Not Set...
2640A/2645A Totalizer Table A-7 provides a summary of the Totalizer specifications. The Totalizer input is located on the DIGITAL I/O connector, terminals Table A-7. 2640A/2645A Totalizer Specification Specification Maximum Input Voltage Minimum Input Voltage Minimum Peak Voltage Isolation Threshold Hysteresis Input Debouncing Maximum Transition Rate Maximum Count...
Users Manual 2640A DC Voltage Measurement Specifications Tables A-9 to A-11 provide 2640A specifications for the dc voltage measurement function. Table A-9. 2640A DC Voltage Measurement General Specifications Specification Input Impedance Normal Mode Rejection Common Mode Rejection Channel-to-Channel Crosstalk Temperature Coefficient Accuracy at -20 C Maximum Input Voltage Table A-10.
Table A-11. 2640A DC Voltage Accuracy Specifications Range 90 Day Slow Fast 90 mV .01%+7 µV .01%+17 µV 300 mV .01%+15 µV .01%+30 µV .01%+.1 mV .01%+.2 mV .01%+1.5 mV .02%+3 mV 150/300V .01%+15 mV .04%+30 mV Note 300V range applies to channels 1 and 11 only. 2640A AC Voltage Measurement Specifications Tables A-12 to A-14 provide 2640A specifications for the ac voltage measurement function.
Users Manual Table A-12. 2640A AC Voltage General Specifications (cont) Specification Maximum Input Voltage Maximum Volt-Hertz Product Temperature Coefficient Accuracy at -20 C DC Component Error Table A-13. 2640A AC Voltage Range and Resolution Specifications Range Full Scale 300 mV 150/300V Note 300V range applies to channels 1 and 11 only.
Table A-14. 2640A AC Voltage Accuracy Specifications Range Frequency 300 mV 20 to 50 Hz 50 to 150 Hz 150 Hz to 10 kHz 10 kHz to 20 kHz 20 kHz to 50 kHz 50 kHz to 100 kHz 20 to 50 Hz 50 to 150 Hz 150 Hz to 10 kHz 10 kHz to 20 kHz...
Users Manual 2640A 4-Wire Resistance Measurement Specifications Tables A-15 to A-17 provide 2640A specifications for the 4-wire resistance measurement function. The 4-wire measurements use 2 input channels a decade apart, e.g., channels 4 and 14. Table A-15. 2640A 4-Wire Resistance Temperature Coefficient Specification Temperature Coefficient Accuracy at -20 C...
2640A 2-Wire Resistance Measurement Specifications The 2640A specifications for the 2-wire resistance measurement function is based on the 4-wire resistance measurement specification (above) except you add a nominal 5-Ohm (10-Ohm maximum) positive offset. This value varies for each channel and with temperature (nominal +1%/ºC). 2640A RTD’s 4-Wire, per ITS-1990 Measurement Specifications Tables A-18 and A-19 provide 2640A specifications for the 4-wire Resistance- Temperature Detector (RTD) measurement function.
Users Manual 2640A RTD’s 2-Wire per ITS-1990 Measurement Specifications The 2640A specifications for the 2-wire Resistance-Temperature Detector (RTD) measurement function is based on the 4-wire RTD measurement specification (above) except you add a nominal 5-Ohm (roughly 13ºC ) positive offset. This value varies for each channel and temperature gradient (nominal +0.4%/ºC).
Table A-21. 2640A Thermocouple Specifications Thermocouple Resolution Type Temperature ºC -100 to 80 80 to 230 230 to 760 -100 to -25 -25 to 120 120 to 800 800 to 1372 -100 to -25 -25 to 120 120 to 1000 1000 to 1300 -100 to -25 -25 to 20...
Users Manual 2640A Frequency Measurement Specifications Tables A-22 to A-23 provide 2640A specifications for the frequency measurement function. Table A-22. 2640A Frequency Accuracy Specifications Frequency Measurement Accuracy, 1 Year, -10ºC to 60ºC Range 15 Hz to 900 Hz 0.01 Hz 900 Hz to 9 kHz 0.1 Hz 9 kHz to 90 kHz...
2645A Specifications This section includes specifications specific to the 2644A instrument by measurement function. 2645A DC Voltage Measurement Specifications Tables A-24 to A-26 provide 2645A specifications for the dc voltage measurement function. Table A-24. 2645A DC Voltage Measurement General Specifications Specification Input Impedance Normal Mode Rejection...
Users Manual Table A-25. 2645A DC Voltage Resolution and Repeatability Specifications Range 90 mV 300 mV 50V/300V* * 300V range applies to channels 1 and 11 only. Table A-26. 2645A DC Voltage Accuracy Specifications Range 90 Day Slow 90 mV .01%+20 µV .01%+50 µV 300 mV...
2645A AC Voltage Measurement Specifications Tables A-27 to A-29 provide 2645A specifications for the ac voltage measurement function. Table A-27. 2645A AC Voltage General Specifications Specification Input Impedance Maximum Crest Factor Crest Factor Error Common Mode Rejection Maximum Input Voltage Maximum Volt-Hertz Product Temperature Coefficient Accuracy at -20 C...
Users Manual Table A-28. 2645A AC Voltage Range and Resolution Specifications Range Full Scale 300 mV Table A-29. 2645A AC Voltage Accuracy Specifications Range Frequency 300 mV 20 to 50 Hz 50 to 150 Hz 150 Hz to 10 kHz 10 kHz to 20 kHz 20 kHz to 50 kHz 50 kHz to 100 kHz...
2645A 4-Wire Resistance Measurement Specifications Tables A-30 to A-32 provide 2645A specifications for the 4-wire resistance measurement function. The 4-wire measurements use 2 input channels a decade apart, e.g., channels 4 and 14. Table A-30. 2645A 4-Wire Resistance Temperature Coefficient Specification Temperature Coefficient Accuracy at -20ºC...
Users Manual 2645A 2-Wire Resistance Measurement Specifications The 2645A specifications for the 2-wire resistance measurement function is based on the 4-wire resistance measurement specification (above) except you add a 700 to 1000 Ohm positive offset. This value varies for each channel and temperature gradient (nominal +1%/ºC).
2645A Thermocouple per ITS-1990 Measurement Specifications A-27. Tables A-35 to A-36 provide 2645A specifications for the thermocouple measurement function per ITS-1990. Table A-35. 2645A Thermocouple General Specifications Specification Input Impedance Open Thermocouple Detect Temperature Coefficient Accuracy at -20 C Table A-36. 2645A Thermocouple Specifications Thermocouple Type Temperature ºC...
Page 234
Users Manual Table A-36. 2645A Thermocouple Specifications (cont) Thermocouple Type Temperature ºC -100 to -25 -25 to 20 20 to 600 600 to 1000 -100 to 0 0 to 150 150 to 400 250 to 600 600 to 1500 1500 to 1767 250 to 1000 1000 to 1400 1400 to 1767...
2645A Frequency Measurement Specifications Tables A-37 to A-38 provide 2645A specifications for the frequency measurement function. Table A-37. 2645A Frequency Accuracy Specifications Frequency Measurement Accuracy, 1 Year, -10ºC to 60ºC Range Slow 15 Hz to 900 Hz 0.01 Hz 900 Hz to 9 kHz 0.1 Hz 9 kHz to 90 kHz 1 Hz...
Noise, Shielding, and Crosstalk Introduction This appendix supplies information on how you can setup your instruments to minimize measurement errors. The topics include static versus dynamic measurement, noise and shielding, and how to minimize crosstalk effects Static Versus Dynamic Measurements Some signals change value slowly.
Page 238
Users Manual reading rate and scan interval that is an exact multiple of the power line period and cause line frequency noise to be aliased to a very low frequency, where it appears as drift, or instability, in the readings. For example, if the instrument is set to Reading Rate = Fast and a scan interval of 100 ms, line frequency noise is aliased because the scan interval is an exact multiple of the period of the power line.
the routing of your input wiring: keep it away from or shielded from all ac power mains wiring and other sources of noise. 3. Channel Spacing When possible, leave as many unused channels as possible between your ac voltage channel inputs and your low level inputs. Crosstalk from a VAC or frequency signal on a 2645A channel to an adjacent channel using the Slow reading rate is as shown in Table B-1: Table B-1.
Users Manual Crosstalk from a VAC or frequency signal on a 2640A channel to an adjacent channel using the slow reading is as shown in Table B-2. VDC 1 k Frequency source <100 Hz -120 dB 100 Hz -120 dB 500 Hz -120 dB 1 kHz...
Introduction The instrument measures the true rms value of ac voltages. In physical terms, the rms (root-mean-square) value of an waveform is the equivalent dc value that causes the same amount of heat to be dissipated in a resistor. True rms measurement greatly implies the analysis of complex ac signals.
Page 242
Users Manual 1.0000V ac (the dc component equals 0). However, consider the 2V (peak-to- peak) square wave. Both types of measurement correctly display the dc component (0V), but the NetDAQ instrument also correctly measures the ac component (1.0000V). The average-responding device measures 1.11V, which amounts to an 11% error. Average-responding measurement devices have been in use for a long time;...
Introduction This appendix discusses Fixed-385 and Custom-385 RTD measurements with NetDAQ. Fixed-385 RTD The Fixed-385 RTD linearization uses segmented polynomials to approximate the Callendar Van-Dusen equation for a 385 RTD. R Logger but , , and 43760 coefficients that conform to ITS90: = 0.00385055 = 1.49978574 = 0.10863383...
Users Manual Custom-385 RTD The Custom-385 RTD linearization solves the Callendar Van-Dusen equation for positive temperatures and uses a polynomial approximation to the Callendar Van- Dusen equation for negative temperatures. In addition to R specified via NetDAQ Logger. nd modified DIN 43760 coefficients that conform to ITS90: = 1.49978574 = 0.10863383 Figure D-1 shows the accuracy envelope for a 385 RTD that has and...
Using the Custom-385 RTD with Other Platinum RTDs Although the Custom-385 RTD linearization uses a other RTD types can be used. The Custom-385 linearization will compensate for the change in , but errors will be introduced due to the difference in the and coefficients of the RTD and the fixed coefficients.
Computed Channel Equations Introduction An equation is converted into a stored binary format which is sent to the instrument where the calculations are performed during the processing of each scan. Constants are passed to the instrument as single precision (4 byte) floating point numbers which have a maximum magnitude of 3.402823E38.
Page 252
Users Manual enclose literal characters or strings which must appear as shown (except for case) () enclose groups of elements to clarify the scope of other symbols <equation> <add-term> <mul-term> <exp-term> <operand> <channel> <constant> <mantissa> <exponent> <function> <function-name> <add-term> { (“+” | “-”) <add-term> } <mul-term { (“*”...
Introduction The ASCII (CSV) data file is maintained in comma-separated ASCII format. The ASCII format is compatible with data analysis and management tools running simultaneously in other computers on a network. A conversion utility, Binary to ASCII Conversion (Utilities menu) is provided for converting data files recorded in Fast Binary (BIN) to ASCII (CSV).
Page 254
Users Manual Data files from an asynchronous instrument contain channel readings from one instrument only. Data files from a group of instruments contain the time tag from the Master instrument and data from all the instruments in the group. Each scan record consists of the instrument time tag and a comma separated list of channel readings (including the DIO word, totalizer, alarm1 and alarm2).
Dynamic Data Exchange (DDE) Introduction Dynamic Data Exchange (DDE) provides NetDAQ Logger data to other Windows applications in real time. For example, with a DDE connection, data collected by NetDAQ Logger can be displayed in a spreadsheet as it is being collected from the instruments.
Page 256
Users Manual DDE reference includes the Service Name, Topics, and Items: (cont) Service Name NETDAQ for 16-bit NETDAQ32 for 32-bit DDE Example Using Excel Complete the following procedure to make a DDE connection between NetDAQ Logger and an Excel spreadsheet. This is presented as an example where instrument BCN=25, is logging data from three configured channels, 13, 14, and 15 into Excel cells A1, A2, and A3 respectively.
Page 257
3. Select cell A1 on the spreadsheet, move the cursor to the Formula Bar and enter the remote reference formula to link cell A1 to instrument 25 channel 13, then <Enter>. (See below.) Select cell A1 Equal Sign Service Name Pipe Character If the instrument is scanning, measurement data is displayed in cell A1.
Page 258
Users Manual 5. Continue by entering the following: Cell 6. Your entries will appear on the spreadsheet as shown below for the conditions entered in this example. The Excel formatting commands allow the Scan Time cell to display as a date/time (for example, 3/4/96, 12:04:00) Formula =NETDAQ32|bcn25!DIO =NETDAQ32|bcn25!Totalizer...
Page 259
Dynamic Data Exchange (DDE) DDE Example Using Excel A note about the Alarm 1 and Alarm 2 values reported in Excel: The alarm value is the decimal equivalent of a 30-bit binary number represents the alarm conditions for each of the 30 instrument channels (20 analog and 10 computed channels).
Page 260
Users Manual 7. Continue with other Excel functions to show charts and features. (See your Excel documentation.) For example, select cells A1, A2, and A3 (click on each cell while holding down the <Cntl> key), then select the New command from the File menu. Click on Chart and then OK.
Introduction This appendix details the unshielded twisted-pair (UTP) cables used for 10BaseT Ethernet interconnection. UTP cable configurations are summarized in Figure H- UTP Cables There are two types of UTP cables: patch cord cables and direct connection cables. Patch cord cables connect an instrument or host computer to an existing 10BaseT hub, usually via a network outlet.
NetDAQ and Network Communication Techniques In the past, Fluke data acquisition instruments depended on RS-232 and IEEE-488 communication methods. This was OK with limited channel measurement rates,...
Page 264
Users Manual low system data throughput because of the processing rate of host computers and disk systems, and instruments that are physically close to the host computer. Now, the increased processing capability of affordable host computers, increased channel measurement rates, and pressure to place the instruments close to the signal source have made support of high-speed, long distance communication techniques necessary.
Page 265
Network Considerations Network Primer 2. Does the network contain more than one subnet? If so, what’s the IP address of my default gateway or router device? 3. Will all the routers route IP packets? 4. What are my domain and host names (optional)? The minimum information you need to get is the IP Addresses for the PC and the instruments.
Page 266
PC. The only trick to 10Base2 wiring is that a terminator must be connected to the cable at each open end. Fluke supplies a terminator with each NetDAQ instrument and wiring kit.
Page 267
A Repeater is a device used to physically isolate and extend the length of a network segment, but it does not divide the network into subnets. A hub is a device that allows communication among multiple 10BaseT network hosts and allows these hosts to communicate with other Ethernet wiring.
Page 268
Users Manual strips off the Frame information around the packet. So, unless you’re doing low- level analysis of network traffic, you never need to deal with Frame information. For any protocol, the Packet consists of a destination address, a source address, some protocol type information, a message length field, and error detection information.
Page 269
The IP address provides two pieces of information: the network ID and the host ID. The network ID must be unique among all network subnets within a connected Internet. It identifies the network subnet to which a host is connected. The host ID must be unique among hosts within a network subnet and identifies the host within the subnet.
Page 270
This is a Class C Net ID assigned to the Fluke Corporation, so it is highly unlikely that it will overlap with a Net ID in use at your company. However, if you want to attach a...
Page 271
software sends out a broadcast frame that every TCP/IP host and router receives. This frame contains a request for the host with the unknown IP address to return a frame to the ARP requester. This response frame contains the Ethernet address of the sourcing host, and this information is then added to the address correspondence table.
Page 272
Users Manual Packet drivers support only one protocol at a time. As a result, only one package of protocol software can use a packet driver at any one time. Packet drivers are usually loaded by executing a *.com or an *.exe file in the autoexec.bat.
Page 273
While a protocol section might look like the following: [NETMANAGE] DRIVERNAME=NETMNG$ BINDINGS=EtherLinkII These two protocol.ini file sections describe a hardware driver characteristic for a 3COM EtherLink II hardware card and specifies that the NetManage protocol stack should bind with the EtherLink II driver. The setup information for ODI drivers is contained in a file called net.cfg that must be in the Novell directory (the driver must be there too).
Page 274
Users Manual Currently, the WINSOCK DLL that operates with LAN Workplace must be downloaded from the Novell forum on CompuServe or you must request a copy from Novell. TCP/IP protocol packages with included WINSOCK DLLs can be purchased from a variety of sources including Microsoft, Banyan, FTP, SunSoft, and IBM.
Page 275
matching network ID. This process is generally referred to as routing the packet, and devices that do this are also referred to as Routers or Gateways. In many modern networks, the use of subnet masking and the default gateway is not necessary.
Page 276
IP address for the PC and instruments. With the isolated network operation, you are limited to using the 198.178.246.0 subnet (Class C subnet assigned to the Fluke Corporation) for both the PC and instruments.
Page 277
Load the network operating system client software first, then load the Trumpet or Newt software. During installation of the Newt software, be careful to enter the IRQ and port address that matches the network card attached to the NetDAQ instruments. I-15...
Page 278
Users Manual Network drivers typically use PC soft interrupts to communicate with the protocol software. If you’re using two network cards, and as a result, two drivers, you must select two different soft interrupt levels for the drivers. For example, Packet drivers (used with Trumpet) usually use soft interrupt level 60 (hex).
Page 279
NetDAQ instrument and the PC default gateway addresses to make sure they are set correctly. Check all network wiring. Make sure that 10Base2 terminators are placed at the two open ends of the coax. Make sure that there is no more than one terminator on each end (some network repeaters, bridges, and routers have termination built into the unit.) In a 10Base2 network, make sure that the collision indicator on the...
Page 280
Users Manual Move your instrument to a less loaded subnet of your network or isolate the NetDAQ instruments on their own network. If possible, increase the scan interval. Use the Fast Binary file format rather than CSV or Trend Link formats. Make sure that there are few or no other Windows applications running at the same time as NetDAQ Logger.
Page 281
7. Failure to communicate when using the Parallel to LAN Adapter (PLA) In some computers the speed of the parallel port hardware is insufficient to support the normal bi-directional communication used by the Silicom PLA. In these rare cases, you must force use of the uni-directional communication on this interface.
Page 282
If this is not the case, one of the recommended Ethernet adapters for your host computer may be purchased from Fluke as an option. See “Installing the Host Computer Ethernet Adapter” before proceeding with the Newt software installation.
Page 283
Network Considerations Newt Networking Software Ethernet Components and Wiring The NetDAQ instruments and host computer should be connected to the network as shown in Chapter 2 of this manual. Instrument Preparation Each NetDAQ instrument should be connected to an ac outlet or dc source with a proper ground.
Page 284
Banyan Vines versions 4.10, 4.11, and 5.5 or higher can use the NDIS driver for network card support. You must configure Vines to use the NDIS driver in order to install the Newt software. This is done via the Banyan PCCONFIG utility supplied with Vines.
Page 285
Network Considerations Newt Networking Software After configuring Vines for NDIS driver support, install Newt. Select "None" when asked for a hardware Vendor. Once Newt is installed, edit the protocol.ini file in the Vines directory to add the following section: [NETMANAGE] DRIVERNAME=NETMNG$ BINDINGS=hardware driver name The hardware driver name should be the same as that used in the drivers section...
Page 286
Users Manual autoexec.bat Netmanage directory added to the path. system.ini netmanage.dos added to the transport= line under the [Network Drivers] section. LoadRMDrivers= will be changed to Yes. protocol.ini [NETMANAGE] section added that should have a BINDINGS= line that links to the adapter driver that is to be used for the interface to the NetDAQ instruments.
Page 287
d. Accept the default directory c:\netmanag. Click Continue. e. Acknowledge the Setup is Complete! message by clicking OK. 3. Select Interface | Add. Accept the Ethernet defaults. Click OK. If you are already using Novell NetWare Client software, Banyan Vines, Windows for Workgroups (with networking enabled) or another package that already has installed an ODI or NDIS driver, the following Hardware Selection dialog box will not appear and the existing driver is used.
Page 288
Users Manual 7. Select Setup | Host Name. Enter the host name assigned by your network administrator for this host computer. Click OK. 8. Select Setup | Domain Name. Enter the domain name assigned by your network administrator for this host computer. Click OK. 9.
Page 289
11. Select File | Save. If the NDIS driver must be loaded, and it cannot be found on one of the directories specified in your PATH environment variable, the following dialog box displays. Enter the path to the NDIS driver. It is usually found in the NDIS directory of the diskette supplied with your Ethernet adapter.
Page 290
Users Manual Write down the Physical (Ethernet) Address from the Custom window inside the back cover of this manual. Select File | Exit to close Newt. I-28...
Newt Error Messages Plug-in Card Failed to get interface 1. No Ethernet card is installed. Complete "Installing the Host characteristics table Computer Ethernet Adapter” in Chapter 2 of this manual. 2. The installed Ethernet card does not match the Newt Custom Failed to initialize Hardware selection.
Users Manual Table I-2. Additions to Host Computer Files made by Newt Installation Newt Installation Newt for a typical plug-in Ethernet card Newt for a typical Parallel-to-LAN Ethernet Adapter I-30 Host Computer Files Addition or Modification Five lines added to c:\autoexec.bat @REM The following line was added by Chameleon (date) C:\NETMANAG/NETBIND C:\NETMANAG;...
Page 293
Network Considerations Glossary Glossary I-24. 10Base2 Network wiring scheme using 50 ohm coaxial cable (10 MHz, Baseband, 200 meter max. length). 10BaseT Also called UTP. Network wiring scheme using unshielded, twisted pair cable (10 MHz, Baseband, Twisted Pair). API Application Programming Interface. A documented interface that allows application software packages to access resources from a library.
Page 294
Domain A group of computers on the network may be assigned a domain name and referred to by that name. Domain names can be used as part of the Internet name of a host. For example, my Internet name is dag@tc.fluke.com. My domain name is “tc”.
Page 295
MAC layer of the network protocol. NetDAQ A trademarked name for a series of really fine instruments, software, and accessories from the Fluke Corporation. Stands for Networked Data Acquisition. NDIS Driver Network Device Interface Specification. One of the three popular device driver specifications used on Personal Computers.
Page 296
Users Manual Packet Driver A network device driver specification defined by 3COM. Packet drivers support operation of only a single protocol stack over the driver and hardware interface. PC Personal Computer. A much misused term usually used to refer to a computer built to comply with the IBM PC defacto standards.
Page 297
change to TCP/IP is contemplated, an RFC is issued over the Internet and everyone is invited to comment on the change before its implementation. Router A router provides a communication path between two or more sections of a network. It directs packets between sections (subnets) of the network so that only those packets addressed to hosts (or other routers) within the subnet are communicated to that subnet.
Page 298
Users Manual UDP User Datagram Protocol. A portion of the TCP/IP protocol. It supports unreliable, connectionless, datagram services. UDP has much lower overhead (both time and space) than TCP, but is unreliable and should be used only when the application software provides algorithms for reliability, sequencing, flow control, and other such services usually provided by TCP.
Error Messages & Exception Introduction This appendix describes error messages and how NetDAQ Logger handles exception conditions during operation. Communications Connection If the communications connection to an instrument is broken during logging, a modeless window appears with a warning message, bringing NetDAQ Logger to the top.
Page 300
Users Manual invalid values into the fields of a dialog box. If a critical instrument error occurs (RAM constants failure or A/D not responding), a modeless window appears with a warning message. For a description of each error message, refer to Error and Status Messages in the online help.
Page 301
Yes to continue operations or No to terminate operations. If the fault recurs, stop operations, and cycle instrument power. If the fault still persists, remove the instrument from service and return it to a Fluke Service Center for repair, along with a description of the error message.
Page 302
Yes to continue operations or No to terminate operations. If the fault recurs, stop operations, and cycle instrument power. If the fault still persists, remove the instrument from service and return it to a Fluke Service Center for repair, along with a description of the error message.
Page 303
Yes to continue operations or No to terminate operations. If the fault recurs, stop operations, and cycle instrument power. If the fault still persists, remove the instrument from service and return it to a Fluke Service Center for repair, along with a description of the error message.
Page 304
Users Manual Associated configuration will change to reflect new instrument model. Continue? You are modifying an instrument model number in the Communications Configuration File, which will also modify the instrument model number for the associated icon on the Icon Bar. Associated icon will change to the new number.
Page 305
Yes to continue operations or No to terminate operations. If the fault recurs, stop operations, and cycle instrument power. If the fault still persists, remove the instrument from service and return it to a Fluke Service Center for repair, along with a description of the error message.
Page 306
Users Manual A Data Set is opened for each instrument in a group instrument and each data set is created by appending 1 of 25 letters in the alphabet. If more than 25 letters are required, then this message appears. To delete old Trend Link files, see "Deleting Old Trend Link Files"...
Page 307
The default location is in the c:\Program Files\Fluke\tl directory. Refer to Chapter 4 of your Trend Link for Fluke manual for more information. Check your Trend Link installation or reinstall Trend Link using the procedures in Chapter 2.
Page 308
Yes to continue operations or No to terminate operations. If the fault recurs, stop operations, and cycle instrument power. If the fault still persists, remove the instrument from service and return it to a Fluke Service Center for repair, along with a description of the error message.
Computed channel [Channel #]: Instrument equation buffer size limits will be exceeded! The instrument’s buffer space for storing equation tokens (1000 bytes) has been exceeded. Delete or modify the equations until they occupy no more than 1000 bytes. If this message continues to appear when you click OK, continue to modify or delete equations as necessary.
Page 310
As long as you do not attempt to set computed channel equations or custom RTDs in NetDAQ Logger, you can continue to operate with your current software. If you want to use these features, you will need to contact your Fluke Service Center for a firmware upgrade.
Page 311
Current Master instrument will become Async. Continue? You are changing an Asynchronous instrument into a Master instrument. A group instrument consists of a Master and one or more Slaves. If you redefine an Asynchronous instrument as a Master, the old Master becomes an Asynchronous instrument.
Page 312
Yes to continue operations or No to terminate operations. If the fault recurs, stop operations, and cycle instrument power. If the fault still persists, remove the instrument from service and return it to a Fluke Service Center for repair, along with a description of the error message.
Page 313
Yes to continue operations or No to terminate operations. If the fault recurs, stop operations, and cycle instrument power. If the fault still persists, remove the instrument from service and return it to a Fluke Service Center for repair, along with a description of the error message.
Page 314
As long as you do not attempt to set computed channel equations or custom RTDs in NetDAQ Logger, you can continue to operate with your current software. If you want to use these features, you will need to contact a Fluke Service Center for a software upgrade.
Page 315
Instrument model does not match! Reconfigure Communications! The instrument model does not match the entry in the Communications Configuration File. Modify the instrument model number in the Communications Configuration File (select the Communications Config command from the Setup menu and click the Modify button).
Page 316
You are attempting to communicate with an instrument whose software is incompatible with this version of NetDAQ Logger. Contact your Fluke Service Center to have your instrument software upgraded. Instruments with invalid BCNs [Setup File Name] not loaded! Configure communications with desired BCNs! You have tried to load instrument icons from a setup file for instruments that do not appear in the Communications Configuration File.
Page 317
Invalid Channel Number! Your equation contains a channel number that is not expressed with the correct syntax. When used within an equation, channel numbers must be preceded by the letter ‘c’ (or ‘C’). The channel number itself is expressed as an integer between 1 and 30.
Page 318
Users Manual Invalid Minimum Value for Scaled Range! You have entered an invalid Minimum Value for the Mx+B Input Range calculation. Enter a Minimum Value that is less than +9.9999E+9. Invalid Minimum Value for Input Range! You have entered an invalid Minimum Value for the Mx+B Input Range calculation.
Page 319
Low on memory! The current Windows memory level cannot support your operations. Close all applications and restart Windows. Maximum number of Instruments reached! You tried to add more than 20 instruments to the Icon Bar. The maximum number of instruments on the Icon Bar is 20. If you wish to add a new instrument, you must first delete an existing instrument on the Icon Bar.
Page 320
You attempted to load a setup file that is not compatible with this version of NetDAQ Logger software. Contact your Fluke Service Center for a software upgrade. You will need to provide a description of this error message along with the current version numbers of both the Setup File and NetDAQ Logger.
Page 321
Yes to continue operations or No to terminate operations. If the fault recurs, stop operations, and cycle instrument power. If the fault still persists, remove the instrument from service and return it to a Fluke Service Center for repair, along with a description of the error message.
Page 322
Users Manual Out of Memory! Close another application and try again! You have run out of Windows memory. Close other applications to free Windows memory and continue operations. PC time earlier than existing Trend Link data! Verify PC time, wait [nn] seconds, or create a new Data Set.
Page 323
Yes to continue or No to terminate operations. If the fault recurs, stop operations, and cycle instrument power. If the fault still persists, remove the instrument from service and return it to a Fluke Service Center for repair, along with a description of the error message.
Page 324
Yes to continue operations or No to terminate operations. If the fault recurs, stop operations, and cycle instrument power. If the fault still persists, remove the instrument from service and return it to a Fluke Service Center for repair, along with a description of the error message.
Page 325
Select at least one trigger type! Your scan parameters do not include a type of trigger. Configure your scan parameters to include at least one of the triggers: Interval Trigger, External Trigger or Alarm Trigger. Selected Trend Link Data Set already contains data for instrument [BCN] Delete the Data Set or create a new Data Set.
Page 326
Users Manual Socket Port Number must range from 1024 to 65535! You tried to enter a Socket Port number less than 1024, greater than 65535 or included extraneous characters. Enter a Socket Port number between 1024 and 65535. The default is 04369. There are no Trend Link data files to be initialized! You do not have Trend Link selected as a data file format for any of the instruments, or instruments with the Trend Link data file format are already...
Page 327
Token String is too Large! This equation exceeds the limit of 1000 bytes after it’s been converted into binary format. Once equations are converted into binary format (so they can be read by the instrument) they cannot exceed 1000 bytes in length. To reduce this equation’s length, reference other computed channels (average, difference, and difference from average) in the equation.
Page 328
Unable to get a Trend Link DLL function address! Your Trend Link installation is missing the required c:\Program Files\Fluke\tl\tbase.dll file. Check that your Trend Link installation is complete. You may also be trying to operate with portions of an older version of Trend Link. Reinstall Trend Link following the procedures in Chapter 2.
Page 329
You tried to open the Communications Configuration File but failed. Check the read/write permissions for the c:\Program Files\fluke\netdaq\ccf.cfg file. This error can be caused by having this file Read-Only. Remove the Read-Only property from the file. Other causes include the ccf.cfg file does not exist or is corrupted. In either of these cases, close the application and delete the current ccf.cfg file.
Page 330
Chapter 3 of this manual.) Unable to read Trend Link default template chart file - tldef.cht! Your NetDAQ Logger installation is missing the required c:\Program Files\Fluke\netdaq\tldef.cht file. The installation of NetDAQ Logger may be corrupted. Reinstall NetDAQ Logger following the procedures in Chapter 2.
Page 331
Yes to continue operations, or click No to terminate operations. If the fault recurs, stop operations and cycle the instrument’s power. If the fault still persists, remove the instrument from service and return it to a Fluke Service Center for repair, along with a description of this error message.
Page 332
Users Manual Window creation failed! Windows was unable to open NetDAQ Logger due to insufficient resources such as insufficient memory in Windows. Close all applications and restart Windows to clear the resource problem. Winsock.dll not found! The winsock.dll file cannot be found in any directory listed in your c:\autoexec.bat path.
Page 339
—1— 10Base2 (Coax) wiring, interconnec- tion, 2-48 10BaseT (twisted-pair) Ethernet wiring , interconnection, 2-48 —2— 2640A 2-Wire resistance measurement specifications, A-13 2640A 4-Wire resistance measurement specifications, A-12 2640A ac voltage measurement specifications, A-9 2640A dc voltage measurement specifications, A-8 2640A frequency measurement specifications, A-16 2640A RTD’s 2-Wire per ITS-1990 measurement specifications, A-14...
Page 340
2640A/2645A NetDAQ Users Manual Accessing NetDAQ Logger commands, 3-3 Accessories, options and, 1-14 Accuracy performance tests, 6-11 Adding an Instrument to the network, 3-5 Adding or Changing the Chart Title, 5-23 Adjusting the Curve Amplitude Scale (Y-axis), 5-16 Adjusting the Curve Time Scale (X-axis), 5-15 Alarm master, 1-10, 2-14...
Page 341
Copying channels, 3-26 Creating instrument icon, 3-8 Crosstalk considerations, 2-11, B-2 Current setup saving the, 3-11 Curve Preferences, 5-18 Curve Status Display, Configuring the, 5-17 Custom-385 RTD, D-2 —D— Data file format, F-1 Data files ASCII, 4-13 binary, 4-13 configuring, 4-14 converting, 4-18 converting from Binary to ASCII, 4-18 trend link, 4-14...
Page 342
2640A/2645A NetDAQ Users Manual Front panel indicators, 2-18 Full-disk, J-1 Fuse replacement, 6-4 —G— General network IP address, reviewing and setting the, 2-37 General network operation, 1-12 General network socket port, reviewing and setting the, 2-36 Glossary, I-31 Grounding and common mode voltage, 2-7 Group instrument operations, 1-13 Group instrument, external trigger wiring for, 2-15...
Page 343
—L— Line frequency, reviewing and setting, 2-29 Linearization, RTD, D-1 Logging status window, 4-7 Logging, simulated, 4-5 Logging, starting and stopping, 4-3 Logging, starting automatically, 3-14 —M— Main window, NetDAQ Logger, 3-3 Maintenance, 6-3 Maintenance of 2640A/2645A instruments, 6-3 Master alarm, 1-10, 2-14 Master alarm output test, 6-29 Master alarm specifications, A-6 Messages, error and status, J-2...
Page 344
2640A/2645A NetDAQ Users Manual Playing Back an ASCII (CSV) File in Trend Link, 5-6 Positioning and rack mounting, 2-5 Power ac, 2-6 dc, 2-7 Power source, connecting to and instrument grounding, 2-5 Power-on options, 2-21 Precision, optimizing performance for, 4-21 Preferences, Background, 5-21 Preferences, Curve, 5-18 Preparation,...
Page 345
ac voltage measurement (2645A), A-19 combined 2640A/2645A, A-1 dc voltage measurement (2640A), A-8 dc voltage measurement (2645A), A-17 frequency measurement (2640A), A-16 frequency measurement (2645A), A-25 general 2640A/2645A, A-1 RTD’s 2-Wire per ITS-1990 measurement specifications (2640A), A-14 RTD’s 4-Wire, per ITS-1990 measurement (2640A), A-13 thermocouple per ITS-1990 measurement (2640A), A-14...
Page 346
2640A/2645A NetDAQ Users Manual Using the Menus, 5-15 Trend Link Chart, Getting the Right Look for your, 5-12 Trend Link for Fluke, Using, 5-3 Trend Link, Installation of, 2-55 Trigger in, A-5 Trigger input, 1-9, 2-13 Trigger input test, 6-30...