2640A/2645A NetDAQ
Users Manual
Custom-385 RTD
The Custom-385 RTD linearization solves the Callendar Van-Dusen equation for
positive temperatures and uses a polynomial approximation to the Callendar Van-
Dusen equation for negative temperatures. In addition to R
specified via NetDAQ Logger. nd
modified DIN 43760 coefficients that conform to ITS90:
= 1.49978574
= 0.10863383
Figure D-1 shows the accuracy envelope for a 385 RTD that has and
match the above coefficients. Because of the error introduced by the linearization
at low temperatures, it is recommended that use of the Custom-385 RTD type be
limited to temperatures between -100 and 600°C.
1.5
1.0
0.5
0.0
-0.5
-1.0
-1.5
-100
D-2
Accuracy Envelope: Custom RTD, 385 Probe
1 Year Slow: 18-28 Degrees C
Alpha: 0.00385055 Delta: 1.49978574 Beta: 0.10863383
2645A
2640A
0
100
Temperature - Degrees C
Figure D-1. 385 RTD
are fixed. These fixed coefficients are the
200
300
400
D-3.
,
may also be
0
that
500
600