Fluke NetDAQ 2640A User Manual page 191

Networked data acquisition unit
Hide thumbs Also See for NetDAQ 2640A:
Table of Contents

Advertisement

Resistance Range*
30 k
30 k
300 k
3 M
* The resistance accuracy in this table makes allowance for up to 0.1 Ohm of lead wire
resistance plus 0.01% decade resistance tolerance. You must add any additional lead wire
resistance present in your setup to the resistance values given in this table.
Resistance Range*
30 k
30 k
300 k
3 M
* The resistance accuracy in this table makes allowance for up to 0.1 Ohm of lead wire
resistance. You must add any additional lead wire resistance present in your setup to the
resistance values given in this table.
5. Close Spy Window.
Decade Resistor
Short Circuit (Zero)
29 k
290 k
2.9 M
5700A
Short Circuit (Zero)
19 k
190 k
1.9 M
Minimum Reading
700
29.681 k
289.07 k
2.8607 M
Minimum Reading
700
19.686 k
189.60 k
1.8740 M
Maintenance
Performance Test
Maximum Reading
1 k
30.019 k
292.63 k
2.9410 M
Maximum Reading
1 k
20.014 k
192.10 k
1.9277 M
6
6-19