4. Verify Accuracy Configure the Decade Resistance Source for the output
values below and verify the Spy window measurement is between the
minimum and maximum values. Change the channel 1 range as required (see
Step 2).
Resistance Range
300
300
3 k
3 k
30 k
300 k
3 M
The resistance accuracy in this table makes allowance for up to 0.01% decade resistance
tolerance.
Resistance Range
300
300
3 k
3 k
30 k
300 k
3 M
5. Close Spy Window.
6-22
Decade Resistor
Short Circuit (Zero)
290
Short Circuit (Zero)
2.9 k
29 k
290 k
2.9 M
5700A
Short Circuit (Zero)
190
Short Circuit (Zero)
1.9 k
19 k
190 k
1.9 M
Minimum Reading
0
289.861
0
2.89849 k
28.9834 k
289.621 k
2.89146 M
Minimum Reading
0
189.912
0
1.89912 k
18.9893 k
189.750 k
1.89425 M
Maximum Reading
0.05
290.139
0.5
2.90137 k
29.0166 k
290.379 k
2.90854 M
Maximum Reading
0.030
190.088
0.30
1.90088 k
19.0107 k
190.250 k
1.90575 M