Wave Generator Verification At 1 M - Fluke 5820A Service Manual

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

Calibrator
Calibrator
Mainframe
Mainframe
Output
Wave Type
(@ 10 kHz)
square
1.8 mV
square
11.9 mV
square
21.9 mV
square
22.0 mV
square
56.0 mV
square
89.9 mV
square
90 mV
square
155 mV
square
219 mV
square
220 mV
square
560 mV
square
899 mV
square
0.90 V
square
3.75 V
square
6.59 V
square
6.6 V
square
30.8 V
square
55.0 V
sine
1.8 mV
sine
21.9 mV
sine
89.9 mV
sine
219 mV
sine
899 mV
sine
6.59 V
sine
55 V
triangle
1.8 mV
triangle
21.9 mV
triangle
89.9 mV
triangle
219 mV
triangle
899 mV
triangle
6.59 V
triangle
55 V
Table 3-24. Wave Generator Verification at 1 M
5790A
Reading
Conversion
(V rms)
5790A Reading x
Conversion Factor
Factor
(V p-p)
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.8284
2.8284
2.8284
2.8284
2.8284
2.8284
2.8284
3.4641
3.4641
3.4641
3.4641
3.4641
3.4641
3.4641
Calibration and Verification
Verification
Tolerance
(V p-p)
0.000154 V
0.000457 V
0.00075 V
0.00076 V
0.00178 V
0.002797 V
0.0028 V
0.00475 V
0.00667 V
0.0067 V
0.0169 V
0.02707 V
0.0271 V
0.1126 V
0.1978 V
0.1981 V
0.9241 V
1.6501 V
0.000154 V
0.000757 V
0.002797 V
0.00667 V
0.02707 V
0.1978 V
1.6501 V
0.000154 V
0.000757 V
0.002797 V
0.00667 V
0.02707 V
0.1978 V
1.6501 V
3
3-51