Ac Current Verification - Fluke 5820A Service Manual

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

Nominal Output
Current @
Frequency
Resistance Value
(Hz)
0.10 mA @ 1k
45
-0.10 mA@ 1k
45
0.10 mA@ 1k
500
-0.10 mA@ 1k
500
0.10 mA@ 1k
1000
-0.10 mA@ 1k
1000
1.099 mA @ 100
45
-1.099 mA @ 100
45
1.099 mA @ 100
500
-1.099 mA @ 100
500
1.099 mA@ 100
1000
-1.099 mA@ 100
1000
1.10 mA@ 100
45
-1.10 mA@ 100
45
1.10 mA@ 100
500
-1.10 mA@ 100
500
1.10 mA@ 100
1000
-1.10 mA@ 100
1000
10.99 mA@ 10
45
-10.99 mA@ 10
45
10.99 mA@ 10
500
-10.99 mA@ 10
500
10.99 mA@ 10
1000
-10.99 mA@ 10
1000
11.00 mA@ 10
45
-11.00 mA@ 10
45
11.00 mA@ 10
500
-11.00 mA@ 10
500
11.00 mA@ 10
1000
Table 3-13. AC Current Verification
Actual
Topline
Baseline
Resistor
Reading
Reading
Value
(Vdc)
0.05% (W)
A
B
Calibration and Verification
Calculated
Current
(Vdc)
(B-C)
A
(mA)
Error
C
D
D - Nom
3
Verification
Tolerance
( mA)
0.000750
0.000750
0.000750
0.000750
0.000750
0.000750
0.003248
0.003248
0.003248
0.003248
0.003248
0.003248
0.003250
0.003250
0.003250
0.003250
0.003250
0.003250
0.027975
0.027975
0.027975
0.027975
0.027975
0.027975
0.028000
0.028000
0.028000
0.028000
0.028000
3-29