Edge Specifications - Fluke 5820A Service Manual

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

1-10. Edge Specifications

The GHz Option offers a Fast Edge function. The specifications for the
Fast Edge function can be found in Chapter 5.
Edge Characteristics into 50
Rise Time
Amplitude Range (p-p)
Resolution
Adjustment Range
Sequence Values
Frequency Range [1]
Typical Jitter, edge to trigger
Leading Edge Aberrations[2]
Typical Duty Cycle
[1] Frequency range above 2 MHz has rise time specification 350 ps typical.
[2] The leading edge aberrations below 250 mV are typical. All readings are referenced to a Tek11801
with an SD26 module or a Tek820 oscilloscope with a 8 GHz bandwidth option.
Note
Table 1-4. Edge Specifications
Load
300 ps
4.0 mV to 2.5 V
4 digits
10 % around each sequence
value (indicated below)
5 mV, 10 mV, 25 mV, 50 mV,
60 mV, 80 mV, 100 mV, 200 mV,
250 mV, 300 mV, 500 mV, 600 mV,
1 V, 2.5 V
1 kHz to 10 MHz
< 3 ps (p-p)
within 2 ns from 50 % of rising edge
2 to 5 ns
5 to 30 ns
after 30 ns
45 % to 55 %
Introduction and Specifications
General Specifications
1-Year Absolute
Uncertainty, tcal
5 C
+0, -100 ps
(2 % of output + 200 V)
(0.33 ppm of setting)
< (3 % of output + 2 mV)
< (2 % of output + 2 mV)
< (1 % of output + 2 mV)
< (0.5 % of output + 2 mV)
1
1-9