Fluke 5820A Service Manual page 19

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

Temperature Coefficient
Relative Humidity
Altitude
Safety
Interface Impedance
Analog Low Isolation
EMC
Line Power
Power Consumption
Dimensions
Weight
Temperature Coefficient for temperatures outside tcal 5 C is 0.1X/ C of
1-year specification.
Operating: < 80 % to 30 C, < 70 % to 40 C,< 40 % to 50 C
Storage: < 95 %, noncondensing
Operating: 3,050 m (10,000 ft) maximum
Nonoperating: 12,200 m (40,000 ft) maximum
Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92
The 5820A is designed to drive both 50
20 V
Complies with EN 61326-1
Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
Line Frequency: 47 to 63 Hz
Line Voltage Variation:
10 % about line voltage setting
250 VA
Height: 17.8 cm (7 inches), standard rack increment, plus 1.5 cm
(0.6 inch) for feet on bottom of unit;
Width: 43.2 cm (17 inches), standard rack width
Depth: 47.3 cm (18.6 inches) overall.
20 kg (44 pounds)
Introduction and Specifications
General Specifications
and 1M
loads.
1
1-7

Advertisement

Table of Contents
loading

Table of Contents