5820A
Service Manual
Nominal Value
(V dc)
0.5
-0.5
1.35
-1.35
2.19
-2.19
2.2
-2.2
6.6
-6.6
10.99
-10.99
11
-11
70.5
-70.5
130
-130
6.599 (50
Calibrator
Mainframe
Output
0 mV
2.49 mV
-2.49 mV
9.90 mV
-9.90 mV
24.9 mV
-24.9 mV
109.9 mV
-109.9 mV
499 mV
-499 mV
2.19 V
-2.19 V
6.599 V
-6.599 V
3-22
Table 3-5. DC Voltage Verification (cont.)
Measured Value
(V dc)
)
Table 3-6. DC Voltage Verification at 50
HP 3458A Rdg (V dc)
Deviation
(V dc)
Reading x Correction
1-Year Spec.
150 V
150 V
362.5 V
362.5 V
572 V
572 V
572 V
572 V
1.67 mV
1.67 mV
2.77 mV
2.77 mV
2.77 mV
2.77 mV
17.65 mV
17.65 mV
32.5 mV
32.5 mV
16.5 mV
Tolerance (V dc)
0.00004 V
4.623E-05 V
4.623E-05 V
6.475E-05 V
6.475E-05 V
0.0001023 V
0.0001023 V
0.0003148 V
0.0003148 V
0.0012875 V
0.0012875 V
0.005515 V
0.005515 V
0.0165375 V
0.0165375 V