This 5820A Service Manual includes: product specifications, appropriate theory of
operation, calibration and verification procedures, maintenance information, and options.

1-6. Specifications

The following paragraphs describe the details for the 5820A specifications. All
specifications are valid after allowing a warm-up period of 30 minutes, or twice the time
the 5820A has been turned off. (For example, if the 5820A has been turned off for 5
minutes, the warm-up period is 10 minutes.)
All specifications apply for the temperature and time period indicated. For temperatures
outside of tcal + 5 C (tcal is the ambient temperature when the 5820A was calibrated),
the temperature coefficient is less than 0.1 times the 1-year specification per C (limited
to 0 C - 50 C).
If you ordered the GHz Option, the following specification tables are supplemented by
the tables with similar headings in Chapter 5:
Edge Specifications
Leveled Sine Wave Specifications
Refer to Figure 1-2 for the dimensional outline of the 5820A Calibrator.