Set the Calibrator Mainframe to the Volt menu. Press  on the Calibrator Mainframe
to activate the output. Then follow these steps to verify ac voltage frequency.
1. Set the PM 6680's FUNCTION to measure frequency on channel A with auto
2. Using the N to BNC cable, connect the CHAN 1 connector on the Calibrator
3. Program the Calibrator Mainframe to output 2.1 V at each frequency listed in Table
4. Allow the PM 6680 reading to stabilize, then record the PM 6680 reading for each
Calibrator Mainframe
Frequency
(Output @ 2.1 V p-p)
10 Hz
100 Hz
1 kHz
10 kHz

3-32. DC Measurement Verification

This procedure uses the following equipment:
1. Select Voltage Measurement mode.
3-26
Figure 3-5. Setup for AC Voltage Frequency Verification
trigger, measurement time set to 1 second or longer, 1M impedance, and filter off.
Mainframe to PM 6680 channel A.
3-9.
frequency listed in Table 3-9. Compare to the tolerance column of Table 3-9.
Table 3-9. AC Voltage Frequency Verification
Fluke 5520A Calibrator
5820A Cable
At 50 MHZ
PM 6680 Reading
(Frequency)
5820A
OSCILLOSCOPE
5820A
CALIBRATOR • 2GHz
CAT
SOURCE/MEASURE
EXT TRIG
CHAN 1
CHAN 2
CHAN 2
130V
PK
MAX
20V PK
SOURCE
MAX
CHAN 3
CHAN 4
30V DC
MAX
MEASURE
CHAN 4
CHAN 1-5
AUX
EXT TRIG
CHAN 5
INPUT
20V PK
20V PK
MAX
MAX
yu057f.eps
Tolerance
-6
3.3 E
-6
33 E
-6
330 E
-6
3,300 E