Fluke 5820A Service Manual page 82

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

5820A
Service Manual
Amplitude
Calibrator
Mainframe
(V)
0.4
0.8
Complete Columns A-E as follows:
A
B
C
D
E
3-44
Table 3-21. High Frequency Flatness Verification (cont.)
A
B
50 kHz
Freq.
(MHz)
50 MHz
100 MHz
150 MHz
200 MHz
250 MHz
300 MHz
350 MHz
400 MHz
450 MHz
500 MHz
550 MHz
600 MHz
50 MHz
100 MHz
150 MHz
200 MHz
250 MHz
300 MHz
350 MHz
400 MHz
450 MHz
500 MHz
550 MHz
600 MHz
Enter the 437B present frequency Reading (W).
Enter the 437B 50 kHz Reading (W).
Apply power sensor correction factor for present frequency (W): CF * (Column A
entry)
Apply power sensor correction factor for 50 kHz (W): CF * (Column B entry)
Compute and enter Error relative to 50 kHz plus floor (%): 100 * [(sqrt(Column
B/Column D entry) -1)
C
D
E
100*(100 V/nominal voltage)].
Calibrator Mainframe
Flatness Spec. (%)
1.52
1.52
2.02
2.02
2.02
2.02
3.52
3.52
3.52
3.52
4.02
4.02
1.51
1.51
2.01
2.01
2.01
2.01
3.51
3.51
3.51
3.51
4.01
4.01