Performance Test and Function Test
Performance Test
Frequency Accuracy Test
This test verifies the E4991A's test signal frequency accuracy with an external frequency
counter.
Specification
Frequency Accuracy:
Required Equipment
Frequency Counter
BNC(m)-BNC(m) Cable, 61 cm
N(m)-BNC(f) Adapter
7mm-N(f) Adapter
Procedure
NOTE
In the following procedure, the 53181A is used. Perform the same operation if another
frequency counter is used.
1. Connect the equipment as shown in Figure 2-1. For testing the analyzer equipped with
NOTE
For testing without Option 1D5, disconnect the above BNC(m)-BNC(m) cable.
28
Description
*1.Opt.050 and Opt.124 can be substituted for Opt.030. In this
case, a N(m)-BNC(f) adapter is necessary.
Option 1D5, connect a BNC(m)-BNC(m) cable between the Ext Ref In connector and
the Ref Oven(Option 1D5) connector on the analyzer rear panel.
10 ppm (without Option 1D5)
1 ppm (with Option 1D5)
Recommended Model
53131/2A Opt.010 and 030
or 53181A Opt.010 and 030
p/n 8120-1839
p/n 1250-0780
11524A
*1
*1