JVC CL-R190 Instruction Manual page 7

19.0-inch (48 cm) color lcd monitor
Table of Contents

Advertisement

Available languages

Available languages

Guidance and manufacturer's declaration – electromagnetic immunity
The model CL-R190 is intended for use in the electromagnetic immunity environment (Home
healthcare environment defined in IEC60601-1-2) specified below. The customer or the user of
the model CL-R190 should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power
frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601 test level
±8 kV contact
±15 kV air
±2 kV for power
supply lines 100 kHz
±1 kV for input/output
lines 100 kHz
±1 kV line(s) to line(s)
±2 kV line(s) to earth
0 % U
(100 % dip in
T
U
)
T
0.5 cycles and
1 cycle
70 % U
(30 % dip in
T
U
)
T
25/30 cycles
0 % U
(100 % dip in
T
U
)
T
250/300 cycles
30 A/m
50Hz or 60Hz
Compliance level
±8 kV contact
±15 kV air
±2 kV for power
supply lines 100 kHz
±1 kV line(s) to line(s)
±2 kV line(s) to earth
0 % U
(100 % dip in
T
U
)
T
0.5 cycles and
1 cycle
70 % U
(30 % dip in
T
U
)
T
25/30 cycles
0 % U
(100 % dip in
T
U
)
T
250/300 cycles
30 A/m
50Hz or 60Hz
- 7 -
Electromagnetic
environment
–guidance
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30 %.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the model
CL-R190 requires
continued operation
during power mains
interruptions, it is
recommended that the
model CL-R190 be
powered from an
uninterruptible power
supply or a battery.
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
The product should be
kept at least 15 cm
away from the source
of power frequency
magnetic fields during
use.
English