JVC CL-R190 Instruction Manual page 10

19.0-inch (48 cm) color lcd monitor
Table of Contents

Advertisement

Available languages

Available languages

portable and mobile RF communications equipment and the model CL-R190
The model CL-R190 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the model CL-R190 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the model CL-R190 as recommended below,
according to the maximum output power of the communications equipment.
Immunity to proximity fields from following RF wireless communication equipment has been
confirmed:
Test
Bandwidth
frequency
(MHz)
(MHz)
385
380-390
450
430-470
710
745
704-787
780
810
871
800-960
930
1720
1845
1700-1990
1970
2450
2400-2570
5240
5500
5100-5800
5785
a):For some services, only the uplink frequencies are included.
b):The carrier shall be modulated using a 50 % duty cycle square wave signal.
English
Recommended separation distances between
a)
a)
Service
Modulation
Pulse
TETRA400
modulation
FM
GMRS 460,
±5 kHz
FRS 460
deviation
1 kHz sine
Pulse
LTE Band
modulation
13, 17
217 Hz
GSM 800 /
900,
TETRA
Pulse
800,
modulation
iDEN 820
18 Hz
CDMA 850,
LTE Band 5
GSM 1800;
CDMA
1900;
Pulse
GSM 1900;
modulation
DECT;
217 Hz
LTE Band 1,
3, 4, 25;
UMTS
Bluetooth,
WLAN,
Pulse
802.11 b/g/
modulation
n,
217 Hz
RFID 2450,
LTE Band 7
Pulse
WLAN
modulation
802.11 a/n
217 Hz
Minimum
Maximum
separation
b)
power
distance
(W)
(m)
1.8
0.3
b)
2
0.3
b)
0.2
0.3
b)
2
0.3
b)
2
0.3
b)
2
0.3
b)
0.2
0.3
- 10 -
IEC/
Compliance
EN60601
level
test level
(V/m)
(V/m)
27
27
28
28
9
9
28
28
28
28
28
28
9
9

Advertisement

Table of Contents
loading