Chapter 8 Tutorial
Relay Life and Preventative Maintenance
Relay Life
As a relay is used, the contacts begin to wear and the resistance of the
closed contacts increases. The initial contact resistance of a relay is
typically 50 m
exceeds 20 to 50 times its initial value, the contact resistance becomes
very erratic and the relay should probably be replaced. For most
applications, a relay with contact resistance greater than 1
replaced. The graph below shows the typical contact resistance
characteristics of the relays used on the 34970A switching modules.
1
0
1k
Relay Load
For most applications, the load switched by the relay is the most important
factor affecting relay life. As shown in the graph below, relay life is
maximized by switching low power. As the power switched increases,
relay life deteriorates.
300V
240V
180V
120V
60V
400
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
(plus lead resistance). When the contact resistance
Typical Relay Lifetime
Full Load
Typical Signals
No Load
10k
100k
1M
10M
Relay Operations (Cycles)
Maximum Switching Capacity
0.2A
0.4A
0.6A
0.8A
1A
10 mA 20 mA 30 mA 40 mA 50 mA
Current Switched
should be
100M
Armature (34901A, 34903A, 34904A, 34908A)
Reed (34902A)