Chapter 8 Tutorial
Measurement Fundamentals
Example: Calculating Measurement Error
Calculate the approximate measurement error for a pulse train input
with a crest factor of 3 and a fundamental frequency of 20 kHz.
The internal DMM is set to the 1 V range. For this example, use the
90-day accuracy specifications of
as shown in chapter 9 on page 406.
= ±(0.05% + 0.04%) = ±0.09%
Error
sine
Error
crest factor
Error
bandwidth
Total Error = 0.09% + 0.15% + 1.4% = 1.6%
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
±
(0.05% of reading + 0.04% of range),
= 0.15%
2
– 3
x 20000
=
*
4 x 3.14159 x 1000000
100 = 1.4%