Agilent Technologies 34901A User Manual page 367

Data acquisition/switch unit
Table of Contents

Advertisement

Chapter 8 Tutorial
Measurement Fundamentals
AC Loading Errors In the ac voltage function, the input of the
internal DMM appears as a 1 M
of capacitance. The cabling that you use to connect signals to the
instrument will also add additional capacitance and loading. The table
below shows the approximate input resistance at various frequencies.
Input Frequency
100 Hz
1 kHz
10 kHz
100 kHz
For low frequencies:
Error (%) =
Additional error for high frequencies:
Error (%) = 100 x
F = Input frequency
R
= Source resistance
s
C
= Input capacitance (150 pF) + Cable capacitance
in
Use low-capacitance cable when measuring high-frequency ac signals
(see page 336).
364
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
resistance in parallel with 150 pF
Input Resistance
700 k
600 k
100 k
10 k
− 100 x R
s
+ 1 M Ω
R
s
é
1
ê
ê
  
ê
1 + (
π
2
x
F x R
ë
s
ù
− 1
ú
ú
ú
)
2
x C
û
in