State Storage Commands
MEMory:STATe
:RECall:AUTO {OFF|ON}
:RECall:AUTO?
Disable or enable (default) the automatic recall of the power-down state
from storage location "0" when power is turned on. Select " ON " to
automatically recall the power-down state when power is turned on.
Select " OFF " to issue a Factory Reset (and state "0" is not automatically
recalled) when power is turned on. The :AUTO? query returns the
automatic recall setting. Returns "0" ( OFF ) or "1" ( ON ).
MEMory:STATe:VALid? {0|1|2|3|4|5}
Query the specified storage location to determine if a valid state is
currently stored in this location. You can use this command before
sending the *RCL command to determine if a state has been previously
stored in this location. Returns "0" if no state has been stored or if it has
been deleted. Returns "1" if a valid state is stored in this location.
MEMory:NSTates?
Query the total number of memory locations available for state storage.
Always returns "6" (memory location "0" is included).