Changing Parameter Settings Via Simatic Pdm; Configuring A New Device; Calibrating Lr460 Via Pdm; Parameters Accessed Via Pull-Down Menus - Siemens SITRANS LR460 HART Operating Instructions Manual

Radar transmitters
Hide thumbs Also See for SITRANS LR460 HART:
Table of Contents

Advertisement

Changing parameter settings via SIMATIC PDM

For a complete list of parameters, see
Note:
1.
Launch SIMATIC PDM, connect to SITRANS LR460, and upload data from the
device.
2.
Adjust parameter values in the parameter value field then press Enter. The status
fields read Changed.
3.
Open the Device menu, click on Download to device, then use File – Save, to save
parameter settings. The status fields are cleared.

Configuring a new device

Configuring a new device
See
newest version of the Device Description (DD).

Calibrating LR460 via PDM

Quick Start Wizard via SIMATIC PDM
See
for a simple application.
For a complete list of parameters see

Parameters accessed via pull-down menus

See Pull-down menus for HART on page 45, and for PROFIBUS on page 46, for a complete
list of parameters that can be accessed only via pull-down menus in SIMATIC PDM.
1

Reset

Factory Defaults
Use Factory Defaults to reset all parameters excluding tag, descriptor, message and
device addresses to the default settings.
1. Open the menu Device – Reset, select Factory Defaults, and click OK.
2. After the reset is complete upload parameters to the PC/PG.
1.
2. 1 .5. Reset
See
A5E42035355
on page 31 to configure your device and load the
on page 49 for a complete list of reset options.
SITRANS LR460 – INSTRUCTION MANUAL
Parameter Reference
on page 31 and follow 4 steps to setup
Parameter Reference
on page 45.
value
fields
on page 45.