9.2 Troubleshooting

Symptom
Power does not turn on.
Measuring temperature is in error.
(out of the specifications)
Noise problem

Table 9-9: Troubleshooting

Probable cause
Power switch is not on.
AC adapter is not connected.
Wrong emissivity is set
Ambient reflection calibration is not made
Focus is not met
High voltage source is nearby.
Connectors or cable touches to any noise
source.
Remedy
Turn on the Power Switch.
Connect AC adapter.
Set correct emissivity.
Execute ambient reflection calibration
Adjust focus correctly.
Take a distance from the high voltage source.
Turn off the power and take a distance from the
noise source.