Water Cooling Enclosure (A-Tv30-Ap-Wc); Figure 8-16: Water Cooling Enclosure With Air Purge; Figure 8-17: Enclosure Connections - Fluke ThermoView TV30 Series User Manual

Thermal imager camera
Table of Contents

Advertisement

®
ThermoView
TV30 Series
Users Manual, Rev. 1.0, May 2023

8.2.3 Water Cooling Enclosure (A-TV30-AP-WC)

The water cooling enclosure allows the camera to be used in high ambient temperatures with water cooling. Water
is supplied at the inlet IN, circulates in the housing walls, cools the unit, and exits at the outlet OUT. The
connections are given via 1/8" NPT fittings.
Max. ambient temperature
The water cooling enclosure comes with an air purge collar. The air purge should be always used to keep the lens
dry.
Air flow rate
Air pressure
Air quality
76
180°C (356°F)
at a flow rate of 2 l / min (0.5 gallon / minute)
and a water inlet temperature of 20°C (68°F)
Chilled water below 10°C (50°F) is not recommended.
30 to 60 l / min (8 to 16 gallon / minute)
max. 5 bar (72.5 psi)
recommended according to ISO8573-1:
class 2 or better for solid particulate
class 4 or better for water
class 2 or better for oil

Figure 8-16: Water Cooling Enclosure with Air Purge

Figure 8-17: Enclosure Connections