The Sleep Mode - Invacare LiNX DLX-REM500 User Manual

Table of Contents

Advertisement

Invacare® LiNX
A
B
User
Function
C
D
E
F
Indirect
Navigation
G
Settings
H
A Enter Rest from drive or seating function via timeout.
B Enter Rest from any User Function via CI.
C Exit Rest via CI configured to enter User Functions.
64
I
Rest
Sleep
J
D Exit Rest via CI specially configured to exit Rest and
return to location before entering Rest.
E Enter Rest from Indirect Navigation via timeout.
F Enter Rest from Indirect Navigation via CI.
G Exit Rest via CI configured to enter Indirect Navigation.
H Exit Rest via CI configured to enter Settings.
I Enter Sleep from Rest via timeout.
J Rest is exited when system is power cycled.

5.17 The sleep mode

The sleep mode is no factory setting, but can be enabled by
your provider. If this parameter is set ON, the system goes
into sleep mode after a period of time without user activity.
This period can be set by the provider.
Before a system goes into sleep mode, the system enters a
transition period. During the transition period, the touch
display and all indicators slowly dim until they are switched
off.
During this transition period sleep mode can be interrupted
by performing any input by moving the joystick, pressing the
ON/OFF button or tapping on the touch display.
To wake the system from sleep mode, move the joystick or
either press the ON/OFF button, if this parameter has been
enabled by your provider.