6 Maintenance

6.1 Replacing Mouthpiece

Risk of Damage to Input Module
Improper mounting of mouthpiece may cause
damage to input module by water or saliva.
– Mouthpiece MUST be completely dry before
installation.
1637425-I
Fig. 6-1
1. Remove mouthpiece A from gooseneck B.
Make sure to leave lipswitch C in shrink sleeving which
keeps together lipswitch and mouthpiece.
2. Insert new mouthpiece.

6.2 Replacing Saliva Trap

Risk of Damage to Input Module
If saliva trap is inserted wrong way round, input
module can get damaged by water or saliva.
– Make sure to insert saliva trap in correct
orientation.
– Saliva trap MUST be installed to reduce risk of
water or saliva entering input module.
1.
Maintenance
Fig. 6-2
115