Routine Maintenance; Draining Liquid From The Filter Housing - Agilent Technologies 4107 User Manual

Nitrogen generator
Table of Contents

Advertisement

Maintenance, Spare Parts and Troubleshooting

Routine Maintenance

The following parts of the nitrogen generator require routine
maintenance.
Daily
Monthly
Yearly

Draining Liquid from the Filter Housing

To drain liquid that has collected in the filter housing:
1
2
Liquid will begin to drain immediately.
26
Check the filter sight glasses for collected liquid and drain if
required
Check power supply and cords for damage
Check gas tubing for leaks.
Confirm the air supply system still complies with all applicable
local regulations
Replace the gas filters
Ensure you have a container to catch the liquid as it begins to
drain to avoid spills.
Loosen the screw at the bottom of the filter housing. See (1) in
the Figure below for the screw location.
Agilent 4107 Nitrogen Generator User's Guide