36.Adjust A2R209 and A2R271 until the start of sweep is aligned to the
leftmost vertical graticule line and the stop sweep is aligned with
the right most vertical graticule line.
37.Press
and
STATE 2
±0.1 divisions.
38.Press
and
STATE 0
±0.1 divisions.
Chapter 2
. The two traces should be aligned within
STATE 3
. The two traces should be aligned within
STATE 1
Adjustment Procedures
2. Display Adjustment (8560E only)
69