Assembly Replacement
Procedure 6A. A6 Power Supply Assembly (8560EC)

Replacement

1. Ensure that the bottom shield wall is in place before replacing the
2. Attach the A6 power supply assembly to the spectrum analyzer
3. Connect W1 to A6J1, W3 to A6J2, fan power wires to A6J3, W8 to
4. Ensure that all cables are safely routed and will not be damaged
5. Secure the power supply cover shield to the power supply using
6. Fold the A2, A3, A4, and A5 assemblies into the spectrum analyzer
Figure 3-16
Power Supply Cover
180
A6 power supply assembly.
chassis and top shield wall using the four screws, torqued to 10-inch
lbs. Attatch all other screws, torqued to 6-inch lbs.
A6J4, and the line-power jack to A6J101. See
183.
when securing the A6 cover.
three flathead screws (1). See
the cover fits into a slot provided in the rear frame assembly. Ensure
that the extended portion of the cover shield is seated in the shield
wall groove.
as described in steps 6 through 12 under "Procedure 5. A2, A3, A4,
and A5 Assemblies Replacement."
Figure 3-17 on page
Figure 3-16 on page
180. One end of
Chapter 3