Replacement; Figure 3-27 A16 And A17 Mounting Screws - HP 8560E Service Manual

Spectrum analyzers
Table of Contents

Advertisement

Figure 3-27
A16 and A17 Mounting Screws

Replacement

1. Connect W7, W8, W9, A6A1W2, and A18W1 to the A17 CRT driver
2. For Option 007 spectrum analyzers: Connect all A16 assembly cables
3. Secure the A17 assembly (and A16 assembly in Option 007) to the
4. Place the spectrum analyzer on its right side frame.
5. Fold the A2, A3, A4, and A5 assemblies into the spectrum analyzer
Chapter 3
assembly. Place the assembly into the center-deck mounting slot
next to the CRT assembly.
as illustrated in
Figure 3-28 on page 202
frame removed so that proper A16 assembly cable routing may be
viewed. Place the A16 assembly into the center-deck mounting slot
nearest the left side frame.
left side frame using two flathead screws (and two spacers in
non-Option 007). For Option 007: Attach the board mounts to the left
side frame using two flathead screws (1). See
201.
as described in "Procedure 5. A2, A3, A4, and A5 Assemblies."
Secure the spectrum analyzer cover assembly.
Procedure 10. A16 FADC/A17 CRT Driver (8560E )
which shows the left side
Figure 3-27 on page
Assembly Replacement
201