Table 2-1
Related Adjustments
Assembly
Changed or
Repaired
A5 IF
A6 Power Supply
A6A1 HV Module
A7 1ST LO
Distribution
Amplifier
A8 Low Band Mixer Frequency Response Adjustment
A9 Input
Attenuator
A10 Tracking
Generator
A11 YTO
A13 2nd Converter
A14 Frequency
Control
A15 RF
Chapter 2
Perform the Following Related Adjustments in the
Order Listed
IF Bandpass Adjustment
IF Amplitude Adjustment
High Voltage Power Supply Adjustment (8560E only)
Display Adjustment (8560E only)
High Voltage Power Supply Adjustment (8560E only)
Display Adjustment (8560E only)
First LO Distribution Amplifier Adjustment
Frequency Response Adjustment (or perform the
Frequency Response Performance Test in the
HP 8560 E-Series and EC-Series Spectrum Analyzer
Calibration Guide. The adjustment must be performed if
the performance test fails.)
Frequency Response Adjustment (or perform the
Frequency Response Performance Test in the
HP 8560 E-Series and EC-Series Spectrum Analyzer
Calibration Guide. The adjustment must be performed if
the performance test fails.)
Tracking Generator Power Level Adjustment
Frequency Response Adjustment
YTO Adjustment
Frequency Response Adjustment (or perform the
Frequency Response Performance Test in the
HP 8560 E-Series and EC-Series Spectrum Analyzer
Calibration Guide. The adjustment must be performed if
the performance test fails.)
Frequency Response Adjustment
Display Adjustment (fast zero span)
YTO Adjustment
First LO Distribution Amplifier Adjustment
Frequency Response Adjustment
10 MHz Reference Adjustment (Option 103)
Calibrator Amplitude Adjustment
External Mixer Bias Adjustment
Sampling Oscillator Adjustment
Signal ID Oscillator Adjustment
External Mixer Amplitude Adjustment
Adjustment Procedures
Introduction
Adjustment
Number
3
4
1
2
1
2
8
10
10
10
9
10
7
10
10
2
7
8
10
12
11
14
6
16
15
53