A4 Log Amp/Cal Oscillator Amplitude Adjustment; Figure 2A-2 If Amplitude Adjustment Locations - HP 8560E Service Manual

Spectrum analyzers
Table of Contents

Advertisement

Figure 2a-2 IF Amplitude Adjustment Locations

A4 Log Amp/Cal Oscillator Amplitude Adjustment

This adjustment sets the output amplitude of the cal oscillator on the
A4 assembly, and the absolute amplitude of the reference 15 dB
attenuator.
The output of the cal oscillator is adjusted so that a −55 dBm signal
applied to the 10.7 MHz IF input on the A5 IF assembly (A5J3) causes a
displayed signal of −60 dBm. The effect of this adjustment is visible
only after the
ADJ CURR IF STATE
output amplitude from the cal oscillator. When the adjustment
sequence is complete, the result of the adjustment should cause the
−35 dBm signal at A5J5 to be displayed at −60 dBm.
This procedure also sets the attenuator of the reference 15 dB
attenuator so that a source amplitude change of 50 dB combined with a
spectrum analyzer reference level change of 50 dB displays an
amplitude difference of 50 dB.
The 15 dB reference attenuator adjustment is preset at the factory and
NOTE
need not be done if the entire A5 IF assembly is replaced.
Chapter 2
Adjustment Procedures: HP 3335A Source not Available
sequence is complete.
ADJ CURR IF STATE
causes the IF gain adjustment to use the "new"
4a. IF Amplitude Adjustments
135

Advertisement

Table of Contents
loading

This manual is also suitable for:

8560ec

Table of Contents