Adjustment Procedures

15. External Mixer Amplitude Adjustment

15. External Mixer Amplitude Adjustment

Assembly Adjusted

A15 RF assembly

Related Performance Test

IF Input Amplitude Accuracy

Description

The slope of the flatness compensation amplifiers is determined. The
user-loaded conversion losses for K-band are recorded and reset to 30
dB. A 310.7 MHz signal is applied to the power sensor and the power
level of the source is adjusted for a −30 dBm reading. The signal is then
applied to the IF INPUT. The flatness compensation amplifiers are then
adjusted (via DACs) to place the displayed signal at the reference level.
Only the determination of the Flatness Compensation Amplifier slope
is performed if the HP 8560E/EC has Option 002.
Figure 2-21
External Mixer Amplitude Adjustment Setup